English
Language : 

U3606A Datasheet, PDF (7/14 Pages) Keysight Technologies – Multimeter|DC Power Supply
Digital multimeter specifications (continued)
Measurement speeds (typical)
Function
Rate
Reading speed1
(readings/second)
Reading speed
over USB2
(readings/second)
DC voltage
Slow (5½ digits)
15
17
(10 V)
Fast (4½ digits)
70
31
DC current
Slow (5½ digits)
15
17
(1 A)
Fast (4½ digits)
70
37
AC voltage
Slow (5½ digits)
15
17
(10 V at 1 kHz)
Fast (4½ digits)
70
31
AC current
Slow (5½ digits)
15
16
(1 A at 1 kHz)
Fast (4½ digits)
70
37
Resistance
Slow (5½ digits)
15
17
(100 kΩ)
Fast (4½ digits)
70
27
Capacitance
Slow/Fast (3½ digits) 5
4.4
(10 µF)
Frequency
Slow (5½ digits)
9
2.7
(voltage path at
Fast (4½ digits)
9
2.7
10 V, 1 kHz)
Frequency
Slow (5½ digit)
9
2.7
(current path at
Fast (4½ digit)
9
2.7
10 V, 1 kHz)
1. Reading rate of the A/D converter.
2. Number of measurements per second that can be read through USB using SCPI “READ?” command.
3. Number of measurements per second that can be read through GPIB using SCPI “READ?” command.
Reading speed
over GPIB3
(readings/second)
17
32
17
36
17
32
17
37
17
31
4.6
2.7
2.7
2.7
2.7
Supplementary specifications
DC voltage
Measurement method
Maximum input voltage
Input impedance
Input protection
Response time
Sigma Delta A-to-D converter
1000 Vdc on all ranges
10 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF
1000 V on all ranges, < 0.3 A short circuit
rms
Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
DC current
Measurement method
Maximum input current
Burden voltage and shunt resistance
Input protection
Response time
Sigma Delta A-to-D converter
10 mA to 3.0 A DC
• < 0.2 V, 10 Ω for 10 mA range
• < 0.2 V, 1 Ω for 100 mA range
• < 0.3 V, 0.1 Ω for 1 A range
• < 0.7 V, 0.01 Ω for 3 A range
Protected with 3.15 A/500 V, FF fuse
Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
7