U3606A Datasheet, PDF (4/14 Pages) Keysight Technologies – Multimeter|DC Power Supply
Digital multimeter specifications
Specification assumptions:
• Specifications stated are after 60-minutes of warm-up and for 5½-digit resolution
• One-year calibration cycle, with calibration temperature of 18 °C to 20 °C
• Operating temperature: 18 °C to 28 °C (64.4 °F to 82.4 °F)
• Accuracy is expressed as ±(% of reading + % of range)
• Temperature coefficient: Add [0.1 x (the specified accuracy) / °C] for 0 °C to 18 °C and 28 °C to 55 °C
• Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for 30 °C to 55 °C
DC specifications
Function Range1
Test current
or burden
voltage
Accuracy ±(% of reading + % of range)
24 hours2 90 days
1 year
23 ºC ± 1 ºC 23 ºC ± 5 ºC 23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
DC voltage
100.000 mV –
1.00000 V
–
10.0000 V
–
100.000 V
–
1000.00 V
–
0.012 + 0.008
0.012 + 0.005
0.012 + 0.005
0.012 + 0.005
0.012 + 0.005
0.015 + 0.008
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.025 + 0.008
0.025 + 0.005
0.025 + 0.005
0.025 + 0.005
0.025 + 0.005
0.0015 + 0.0005
0.0010 + 0.0005
0.0020 + 0.0005
0.0015 + 0.0005
0.0015 + 0.0005
DC current3
10.0000 mA
100.000 mA
1.00000 A
3.0000 A
< 0.2 V
< 0.2 V
< 0.3 V
< 0.7 V
0.05 + 0.015
0.05 + 0.005
0.05 + 0.007
0.05 + 0.007
0.05 + 0.015
0.05 + 0.005
0.05 + 0.007
0.05 + 0.007
0.05 + 0.015
0.05 + 0.005
0.15 + 0.007
0.15 + 0.007
0.0060 + 0.0005
0.0060 + 0.0005
0.0100 + 0.0005
0.0150 + 0.0010
Resistance4
100.000 Ω
1000.00 Ω
10.0000 kΩ
100.000 kΩ
1.00000 MΩ
10.0000 MΩ
100.000 MΩ
0.83 mA
0.83 mA
100 µA
10 µA
900 nA
205 nA
205 nA || 10 MΩ
0.04 + 0.008
0.04 + 0.005
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.20 + 0.005
1.60 + 0.005
0.04 + 0.008
0.04 + 0.005
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.20 + 0.005
1.60 + 0.005
0.05 + 0.008
0.05 + 0.005
0.05 + 0.005
0.05 + 0.005
0.06 + 0.005
0.25 + 0.005
2.00 + 0.005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0150 + 0.0005
0.1500 + 0.0005
Continuity
1.0000 kΩ
0.83 mA
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.0050 + 0.0005
Diode5
1.0000 V
0.83 mA
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.0050 + 0.0005
Capacitance6 1.000 nF
0.75 µA
–
–
2.0 + 0.8
0.02 + 0.001
10.00 nF
0.75 µA
–
–
1.0 + 0.5
0.02 + 0.001
100.00 nF
8.3 µA
–
–
1.0 + 0.5
0.02 + 0.001
1.000 µF
83 µA
–
–
1.0 + 0.5
0.02 + 0.001
10.00 µF
83 µA
–
–
1.0 + 0.5
0.02 + 0.001
100.0 µF
83 µA
–
–
1.0 + 0.5
0.02 + 0.001
1000 µF
0.83 mA
–
–
1.0 + 0.5
0.02 + 0.001
10000 µF
0.83 mA
–
–
2.0 + 0.5
0.02 + 0.001
1. 20% over-range on all ranges, except for 1000 V range.
dc
2. Relative to calibration standards.
3. Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a lower current or offset current im-
mediately after a high-current measurement, ensure that the U3606A has cooled down.
4. Specifications stated are for 2-wire resistance measurements using Null math operation. Without Null, add a 0.2 Ω error. To eliminate noise interfer-
ence which may be induced by the test leads, a shielded test cable is recommended for resistances above 100 kΩ.
5. Specifications stated are for the voltage measured at the input terminals only. The test current (1 mA) is typical. Variation in the current source will
create some variation in the voltage dropped across a diode junction.
6. Specifications stated are for open test lead measurements with film capacitor or better, using Null math operation.
4