English
Language : 

U3606A Datasheet, PDF (5/14 Pages) Keysight Technologies – Multimeter|DC Power Supply
Digital multimeter specifications (continued)
Low-resistance specifications
Accuracy ±(% of reading + % of range)1
Range
Test current
1 year (23 ºC ± 5 ºC)
100 mΩ
1.0000 A
0.25 + 0.05
1000 mΩ
0.1000 A
0.25 + 0.03
10 Ω
0.1000 A
0.25 + 0.03
1. Four-wire measurement method is used. Test current is sent from the FORCE terminals and resistance is measured at the SENSE terminals.
AC specifications
Accuracy ±(% of reading + % of range)
Function Range1
Frequency
range
24 hours2 90 days
1 year
23 ºC ± 1 ºC 23 ºC ± 5 ºC 23 ºC ± 5 ºC
True rms AC
voltage3
100.000 mV
to 750.00 V4
20 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 30 kHz
30 kHz to 100 kHz7
0.60 + 0.1
0.16 + 0.1
0.80 + 0.1
3.00 + 0.2
0.60 + 0.1
0.16 + 0.1
0.80 + 0.1
3.00 + 0.2
1.00 + 0.15
0.20 + 0.1
1.00 + 0.16
3.00 + 0.28, 9
True rms AC
current
10.0000 mA
to 3.0000 A10
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz11
0.80 + 0.1
0.40 + 0.1
2.00 + 0.2
0.80 + 0.1
0.40 + 0.1
2.00 + 0.2
1.50 + 0.1
0.50 + 0.1
2.00 + 0.2
1. 20% over-range on all ranges, except for 750 V range.
ac
2. Relative to calibration standards.
3. Specifications stated are for input signals greater than 5% of range.
4. Available ranges: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V.
5. For 750 V range accuracy is specified for input less than 200 V .
rms
6. For 100 mV range, the 10 kHz to 30 kHz accuracy is 1.5 + 0.3.
7. Additional error 0.003% of full scale per kHz to be added when signal input changes less than 10% of range.
8. For 100 mV range, the 30 kHz to 100 kHz accuracy is 5 + 0.3.
9. For 750 V range accuracy is specified for input less than 300 V .
rms
10. Available ranges: 10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A.
11. For 1 A and 3 A ranges, the accuracy is specified for frequencies less than 5 kHz.
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
0.05 + 0.02
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
5