English
Language : 

U3606A Datasheet, PDF (12/14 Pages) Keysight Technologies – Multimeter|DC Power Supply
General characteristics
Power supply
Power consumption
Current Input Fuse
Display
Operating environment
Storage compliance
Safety compliance
EMC compliance
Shock and vibration
Remote interface
Programming language
Measurement category
Dimensions (H × W × D)
Weight
Warranty
Calibration cycle
Warm-up time
•
Universal 100 V to 240 V ±10%
ac
ac
• AC line frequency 45 Hz to 66 Hz (360 Hz to 440 Hz for 100/120 V operation)
150 VA maximum
3.15 A, 500 V FF fuse (on front panel)
Highly visible vacuum-fluorescent display (VFD)
• Operating temperature from 0 °C to +55 °C
• Relative humidity up to 80% at 30 °C RH (non-condensing)
• Altitude up to 2000 meters
• Pollution degree 2
• For indoor use only
–40 °C to 70 °C
Certified with:
• IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
• Canada: CAN/CSA-C22.2 No. 61010-1-04
• USA: ANSI/UL 61010-1:2004
Certified with:
• IEC61326-1:2005 / EN61326-1:2006
• CISPR 11:2003 / EN55011:2007 (Group 1 Class A)
• Canada: ICES/NMB-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
Tested to IEC/EN 60068-2
• GPIB IEEE-488
• Full Speed USB 2.0 (Type B)
• USBTMC 488.2 Class device
• USB-CDC
Standard Commands for Programmable Instruments (SCPI)
• CAT II, 300 V
•
CAT I, 1000 V , 750 V rms
dc
ac
• 2500 Vpk transient over voltages
105 × 255 × 329 mm (with rubber bumpers)
87 x 215 x 312 mm (without rubber bumpers)
3.775 kg (with rubber bumpers)
3.535 kg (without rubber bumpers)
Three years
One year
60 minutes
12