English
Language : 

705-5100-121 Datasheet, PDF (47/56 Pages) Magnetrol International, Inc. – GuidedWave Radar Level Transmitter
7.3.2 Performance - Model 705
Reference Conditions Œ
Linearity 
Coaxial/Twin Rod Probes:
Single Rod Probes:
Measured Error Coaxial/Twin Rod Probes:
Single Rod Probes
Resolution
Repeatability
Hysteresis
Response Time
Warm-up Time
Operating Temp. Range
LCD Temp. Range
Ambient Temp. Effect
Process Dielectric Effect
Humidity
Electromagnetic Compatibility
Reflection from water at +70° F (+20° C) with 72" coaxial probe
(CFD threshold)
<0.1% of probe length or 0.1 inch (whichever is greater)
<0.3% of probe length or 0.3 inch (whichever is greater)
±0.1% probe length or 0.1 inch (whichever is greater)
±0.5% probe length or 0.5 inch (whichever is greater)
±0.1 inch
<0.1 inch
<0.1 inch
<1 second
<5 seconds
-40° to +175° F (-40° to +80° C)
-5° to +160° F (-20° to +70° C)
Approximately +0.02% of probe length/ ° C
<0.3 inch within selected range
0-99%, non-condensing
Meets CE requirements (EN-61000-6-2/2001, 61000-6-4/2001)
(Twin and Single Rod probes must be used in metallic vessel or
stillwell to maintain CE requirement)
ΠSpecifications will degrade with Model 7xB, 7xD, and 7xP probes and/or fixed threshold configuration.
 Top 24 inches of Model 7xB probe: 1.2 inches (30 mm). Specification for top 48 inches of single rod will
be application dependent.
7.3.3 Performance - Model 705 Interface
Reference Conditions
Linearity
Measured Error
Upper Layer Dielectric
Interface Layer Dielectric
Resolution
Repeatability
Hysteresis
Response Time
Warm-up Time
Operating Temp. Range
LCD Temp. Range
Ambient Temp. Effect
Humidity
Electromagnetic Compatibility
Reflection from liquid of selected dielectric at +70° F (+20° C) with 72" probe
<0.5 inch
Upper layer ±1 inch
Interface layer ±1 inch (clean distinct interface required)
1.4–5.0
>15
±0.1 inch
<0.5 inch
<0.5 inch
<1 second
<5 seconds
-40° to +175° F (-40° to +80° C)
-5° to +160° F (-20° to +70° C)
Approximately ±0.02% of probe length/ ° C
0-99%, non-condensing
Meets CE requirements (EN-61000-6-2/2001, EN 61000-6-4/2001)
(Twin and Single Rod probes must be used in metallic vessel or
stillwell to maintain CE requirement)
57-640 Eclipse Guided Wave Radar Transmitter - FOUNDATION fieldbus