705-5100-110 Datasheet, PDF (32/68 Pages) Magnetrol International, Inc. – Enhanced Model 705
2.6.5.4 Measurement Type: Interface and Volume
Display
45
Hist Rst
46
FidTicks
xxxx
Action
Press Enter and select yes
to clear history
Diagnostic Display
Comment
Similar to SZ Alarm Reset
Time of flight from start of ramp to fiducial
47
Fid Sprd
48
Fid Type
(select)
Superuser Parameter
49
Fid Gain
xxx
Superuser Parameter
50
Window
xxx
Factory Parameter
51
Conv Fct
xxxx
Factory Parameter
52
Scl Ofst
xxx
Factory Parameter
53
Neg Ampl
xxx
Superuser Parameter
54
Ifc Ampl
Superuser Parameter
55
Pos Ampl
xxx
Superuser Parameter
56
Signal
xxx
Diagnostic Display
57
Compsate
(select0
Superuser Parameter
58
7xKCorr
xxx
Superuser Parameter
59
ElecTemp
xxx C
Diagnostic Display
60
Max Temp
xxx C
Superuser Password
61
Min Temp
xxx C
Superuser Password
62
SZ Hyst
xx.x lu
Superuser Password
Select from positive or negative
(Selection only allowed for some probes, fixed for others)
Calibration parameter
Calibration parameter
Indication of level signal amplitude
Select from None, Manual, Auto
Distance in mm from fiducial to user reference point
(7xK probe characteristic)
Present temperature in electronics compartment
(degrees Celsius)
Maximum electronics temperature recorded
Minimum electronics temperature recorded
Diagnostic factory setting
32
57-600 Eclipse Guided Wave Radar Transmitter