English
Language : 

705-5100-110 Datasheet, PDF (31/68 Pages) Magnetrol International, Inc. – Enhanced Model 705
2.6.5.4 Measurement Type: Interface and Volume
Display
24
BlockDis
xx.x lu
Action
Comment
Enter distance below
Allows user to ignore level measurements near the top of the
reference point where level is probe
not sensed
Select lop current behavior Safety Zone is a user-defined area just below the Blocking
when level is sensed in safety Distance. Enable Fault if necessary to ensure safe, reliable high-
25
SZ Fault
(select)
zone
level readings in critical applications. Choices are None, 3.6 mA,
22 mA, Latch 3.6 or Latch 22. If Latch 3.6 or Latch 22 is
selected, the loop current will remain in alarm until it is manually
cleared with the SZ Alarm Reset below (#27)
Enter distance below
Enter a distance value that develops a safety zone just below the
26
SZ Height
xx.x lu
BlockDis where SZ Fault will Blocking Distance. Here the unit will report a Safety Zone Fault
be asserted
(#25) if the level rises into this area.
27
Press Enter to clear a
SZ Alarm Reset
latched Safety Zone alarm
Clear a latched Safety Zone alarm
28
Threshld
(select)
Select from CFD, Fixed
For interface, refers to threshold for upper level pulse. (Set to
Fixed for most common applications.)
29
IfcThrsh
(select)
Select from CFD, Fixed
Interface mode only. Threshold for interface level pulse. (Set to
CFD for most common applications.)
30
Poll Adr
xx
Enter HART polling address
number (0-15)
31
Trim Lvl
xx.x lu
Enter value to adjust Level -10.0 inches <= Lvl Trim <= +10.0 inches
reading
(Requires superuser password)
32
Trim 4
xxxx
Fine tune the 4 mA point
Adjust setting to output exactly 4.0 mA on current meter
33
Trim 20
xxxx
Fine tune the 20 mA point Adjust setting to output exactly 20.0 mA on current meter
34
Loop Tst
xx.x mA
Enter a mA Output value
Set mA output to any given value to perform loop test
35
LvlTicks
Xxxxx
Diagnostic Display
Time of flight from fiducial to interface signal
36
IfcTicks
Xxxxx
Diagnostic Display
Interface mode only
Time of flight through upper liquid
37
Medium
Diagnostic Display
Interface mode only
Displayed messages are: Unknown, Oil Only, Thin Oil, Thick Oil
38
New Pass
Xxx
Enter new password (0-255) Displays encrypted value of present password
39
Language
(select)
Select from English, Spanish, Language choice for LCD display. (no HART counterpart)
French, German
40
Mdl705HT
Ver3.0a0
Transmitter display
Product identification. Firmware version
41
DispFact
(select)
Select Yes to display factory Allows for viewing the factory parameters
parameter menus
42
History
Press Enter to view history Diagnostic Display
(current status) of recent exceptions
43
HF cable
(select)
Superuser Parameter
Select from 3-foot or 12-foot remote
44
Run Time
57-600 Eclipse Guided Wave Radar Transmitter
31