STEP 7: FALT (Fault)
1. FALT (fault) refers to the choice of the alarm relay to DNRG
(de-energize) or IGNR (ignore) upon diagnosing a fault.
NOTE: The dedicated diagnostic Relay 1 will always de-energize
when a fault is detected.
When FALT is displayed, press ENT ENT to display DNRG.
2. If this is the desired condition, press ENT ENT to display
CAL.
3. If IGNR is desired, use the UP or DOWN
arrow key to toggle between the selections.
4. When the appropriate selection is displayed, press ENT ENT
to lock in the value and display CAL.
STEP 8: CAL (Calibration)
1. After modifying any or all of the preceding values, CAL
will lock in all the values and store them in memory.
Press ENT ENT to display NO (do not calibrate).
2. Press ENT ENT if it is desired to leave current values.
3. If new values are desired, use the UP and DOWN
arrow keys to toggle to YES (calibrate).
4. Press ENT ENT to lock all the new values in memory and
return to the RUN mode.
NOTE: Ensure that the level of the process media is at the exact point
that corresponds to the choice made in Step 1 LEVL:DRY (not
covered with media) or WET (covered with media). This is crit-
ical to the proper calibration of the instrument. The unit will
take a new capacitance reading (establish a new set point)
when YES is chosen and ENT ENT is depressed.
15
50-621 Kotron Sentinel Models 822, 832 and 842 RF Point Level Switches