E4980A-1CM Datasheet, PDF (7/38 Pages) Keysight Technologies – Precision LCR Meter
07 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet
Absolute measurement accuracy
The following equations are used to calculate absolute accuracy.
Absolute accuracy Aa of |Z|, |Y|, L, C, R, X, G, B (L, C, X, and B accuracies apply when
Dx ≤ 0.1, R and G accuracies apply when Qx ≤ 0.1 )
Equation 1.
Aa = Ae + Acal
Aa
Absolute accuracy (% of reading value)
Ae
Relative accuracy (% of reading value)
Acal Calibration accuracy (%)
where G accuracy is applied only to G-B measurements.
D accuracy (when Dx ≤ 0.1)
Equation 2.
De + θcal
Dx
Measured D value
De
Relative accuracy of D
θcal Calibration accuracy of θ (radian)
Q accuracy (When Qx × Da < 1)
Equation 3.
(Qx2 × Da)
± —(—1—±———Q—x—×——D—a—)
Qx
Measured Q value
Da
Absolute accuracy of D
θ accuracy
Equation 4.
θe + θcal
θe
Relative accuracy of θ (degree)
θcal Calibration accuracy of θ (degree)