English
Language : 

E4980A-1CM Datasheet, PDF (14/38 Pages) Keysight Technologies – Precision LCR Meter
14 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet
Effect of cable length
Table 21. When the cable is extended, multiply Yo by the following factor.
Test
Cable length
frequency [Hz] 0 m 1 m
2m
4m
100 - 100 k
1
1 + 5 × Fm/1 MHz 1 + 10 × Fm/1 MHz 1 + 20 × Fm/1 MHz
100 k - 1 M
1
1 + 0.5 × Fm/1 MHz 1 + 1 × Fm/1 MHz 1 + 2 × Fm/1 MHz
1M-2M
1
1 + 1 × Fm/1 MHz 1 + 2 × Fm/1 MHz 1 + 4 × Fm/1 MHz
Fm [Hz] Test frequency
Temperature factor Kt
Table 22. The temperature factor Kt is given below.
Temperature [°C]
Kt
0 - 18
4
18 - 28
1
28 - 55
4