E4980A-1CM Datasheet, PDF (31/38 Pages) Keysight Technologies – Precision LCR Meter
31 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet
Measurement time
Definition
This is the time between the trigger and the end of measurement (EOM) output on the
handler interface.
Conditions
Table 54 shows the measurement time when the following conditions are satisfied:
–– Normal impedance measurement other than Ls-Rdc, Lp-Rdc, Vdc-Idc
–– Impedance range mode: hold range mode
–– DC bias voltage level monitor: OFF
–– DC bias current level monitor: OFF
–– Trigger delay: 0 s
–– Step delay: 0 s
–– Calibration data: OFF
–– Display mode: blank
Table 54. E4980A measurement time [ms](DC bias:OFF)
Measurement Test frequency
time mode
20 Hz 100 Hz 1 kHz
10 kHz
1 LONG
480
300
240
230
2 MED
380
180
110
92
3 SHORT
330
100
20
7.7
100 kHz
220
89
5.7
1 MHz
220
88
5.6
2 MHz
220
88
5.6
10
1
1. LONG
2. MED
3. SHORT
0.1
0.01
0.001
20
100
1k
10k
100k
Test frequency [Hz]
Figure 9. Measurement time (E4980A, DC bias: OFF)
1M 2M