Comparator function of list sweep: The comparator function enables setting one pair of
lower and upper limits for each measurement point.
You can select from: Judge with the first sweep parameter/Judge with the second
parameter/Not used for each pair of limits.
Time stamp function: In the sequential mode, it is possible to record the measurement
starting time at each measurement point by defining the time when FW detects a trigger
as 0 and obtain it later with the SCPI command.
Comparator function
Bin sort: The primary parameter can be sorted into 9 BINs, OUT_OF_BINS, AUX_BIN,
and LOW_C_REJECT. The secondary parameter can be sorted into HIGH, IN, and LOW.
The sequential mode and tolerance mode can be selected as the sorting mode.
Limit setup: Absolute value, deviation value, and % deviation value can be used for setup.
BIN count: Countable from 0 to 999999.
DC bias signal
Table 29. Test signal voltage
Range
0 V to +2 V
Resolution
0 V / 1.5 V / 2 V only
Accuracy
0.1% + 2 mV (23 °C ± 5 °C)
(0.1% + 2 mV) × 4
(0 to 18 °C or 28 to 55 °C)
Output impedance:
100 Ω (nominal)
Note
The following USB memory can be used.
Complies with USB 1.1; mass storage class,
FAT16/FAT32 format; maximum consumption
current is below 500 mA.
Recommended USB memory: 4 GB USB Flash
memory (Keysight PN 1819-0637).
Use the recommended USB memory device
exclusively for the E4980A, otherwise,
previously saved data may be cleared.
If you use a USB memory other than the
recommended device, data may not be
saved or recalled normally.
Keysight will NOT be responsible for data loss in
the USB memory caused by using the E4980A.
Measurement assistance functions
Data buffer function: Up to 201 measurement results can be read out in a batch.
Save/Recall function:
–– Up to 10 setup conditions can be written to/read from the built-in non-volatile
memory.
–– Up to 10 setup conditions can be written to/read from the USB memory.
–– Auto recall function can be performed when the setting conditions are written to
Register 10 of the USB memory.
Key lock function: The front panel keys can be locked.
GPIB: 24-pin D-Sub (Type D-24), female; complies with IEEE488.1, 2 and SCPI
USB host port: Universal serial bus jack, type-A (4 contact positions, contact 1 is on
your left), female (for connection to USB memory only).
USB interface port: Universal serial bus jack, type mini-B (4 contact positions); complies
with USBTMC-USB488 and USB 2.0; female; for connection to the external controller.
USBTMC: Abbreviation for USB Test & Measurement Class
LAN: 10/100 BaseT Ethernet, 8 pins (two speed options)
LXI Compliance: Class C (only applies to units with firmware revision A.02.00 or later)