|
E4980A-1CM Datasheet, PDF (11/38 Pages) Keysight Technologies – Precision LCR Meter | |||
|
◁ |
11 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz â Data Sheet
Example of C-D accuracy calculation
Measurement conditions
Test Frequency:
Measured C value:
Test signal voltage:
Measurement time mode:
Measurement temperature:
1 kHz
100 nF
1 Vrms
Medium
23 °C
Ab = 0.05%
|Zm| = 1 / (2Ï Ã 1 à 103 à 100 à 10-9) = 1590 Ω
Zs = 0.6 mΩ à (1 + 0.400/1) à (1 + â(1000/1000) = 1.68 mΩ
Yo = 0.5 nS Ã (1 + 0.100/1) Ã (1 + â(100/1000) = 0.72 nS
C accuracy: Ae = [0.05 + 1.68 m/1590 à 100 + 0.72 n à 1590 à 100] à 1 = 0.05%
D accuracy: De = 0.05/100 = 0.0005
Basic accuracy
Basic accuracy Ab is given below.
Table 12. Measurement time mode = SHORT
Test
frequency [Hz]
5 mVrms -
50 mVrms
Test signal voltage
50 mVrms - 0.3 Vrms -
0.3 Vrms 1 Vrms
20 - 125
(0.6%) Ã
(50 mVrms/Vs)
0.60%
0.30%
125 - 1 M
(0.2%) Ã
(50 mVrms/Vs)
0.20%
0.10%
1M-2M
(0.4%) Ã
(50 mVrms/Vs)
0.40%
0.20%
1 Vrms -
10 Vrms
0.30%
0.15%
0.30%
10 Vrms -
20 Vrms
0.30%
0.15%
0.30%
Table 13. Measurement time mode = MED, LONG
Test
frequency [Hz]
Test signal voltage
5 mVrms - 50
mVrms
50 mVrms -
0.3 Vrms
0.3 Vrms -
1 Vrms
20 - 100
(0.25%) Ã
(30 mVrms/Vs)
0.25%
0.10%
100 - 1 M
(0.1%) Ã
(30 mVrms/Vs)
0.10%
0.05%
1M-2M
(0.2%) Ã
(30 mVrms/Vs)
0.20%
0.10%
1 Vrms -
10 Vrms
0.15%
0.10%
0.20%
10 Vrms -
20 Vrms
0.15%
0.15%
0.30%
Vs [Vrms]
Test signal voltage
|
▷ |