English
Language : 

E5072A Datasheet, PDF (6/23 Pages) Keysight Technologies – E5072A ENA Series Network Analyzer
6 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet
Corrected system performance with type-N device connectors,
85092C electronic calibration (ECal) module
Network analyzer : E5072A
Calibration kit : 85092C (Type-N, 50 Ω) Electronic calibration (ECal) module
Calibration
: full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C )
with < 1 °C deviation from calibration temperature, isolation calibration is not performed.
Description
Directivity
Source match
Load match
Reflection tracking
Transmission tracking
Specification (dB)
300 kHz to 10 MHz
45
36
41
± 0.100
± 0.053
10 MHz to 3 GHz
54
44
47
± 0.040
± 0.040
3 to 6 GHz
52
41
44
± 0.060
± 0.069
6 to 8.5 GHz
47
36
39
± 0.070
± 0.136
Magnitude
10
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
1
6 GHz to 8.5 GHz
0.1
0.01
10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90
Transmission coefficient(dB)
Figure 4. Transmission uncertainty (Specification)
Magnitude
0.05
0.04
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
0.03
6 GHz to 8.5 GHz
0.02
0.01
0
0
0.2
0.4
0.6
0.8
1
Reflection coefficient(linear)
Figure 5. Reflection uncertainty (Specification)
Phase
100
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
10
6 GHz to 8.5 GHz
1
0.1
10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90
Transmission coefficient(dB)
Phase
10
8
6
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 8.5 GHz
4
2
0
0
0.2
0.4
0.6
0.8
1
Reflection coefficient(linear)