English
Language : 

E5072A Datasheet, PDF (12/23 Pages) Keysight Technologies – E5072A ENA Series Network Analyzer
12 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet
Stability
Description
Magnitude
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz
Phase
30 to 300 kHz
300 kHz to 3 GHz
3 to 6 GHz
6 to 8.5 GHz
Specification
SPD
± 0.04 dB/°C
± 0.005 dB/°C
± 0.03 dB/°C
± 0.04 dB/°C
± 0.8 deg/°C
± 0.1 deg/°C
± 0.4 deg/°C
± 0.8 deg/°C
Dynamic Accuracy 1
Description
Specification (dB)
Typical
Magnitude
10 dBm
–30 dBm
–100 dBm
–110 dBm
± 0.21 dB
± 0.045 dB
± 2 dB
± 3.0 dB
Phase
10 dBm
–30 dBm
–100 dBm
± 5 deg
± 0.3 deg
± 15 deg
1.  Accuracy of the test port input power reading is relative to –10 dBm reference input power level.
Magnitude
10
Phase
100
1
0.1
0.01
10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100
Testport Power (dBm)
Figure 10. Dynamic Accuracy
10
1
0.1
10 0 -10 -20 -30 -40 -50 -60 -70 -80 -90 -100
Testport Power (dBm)