English
Language : 

E5072A Datasheet, PDF (15/23 Pages) Keysight Technologies – E5072A ENA Series Network Analyzer
15 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet
General Information
Measurement Receiver Input (R1, R2, A, B)
Description
System band width
Range
General characteristic
10 Hz to 500 kHz
Nominal settings are:
10, 15, 20, 30, 40, 50, 70, 100, 150, 200, 300, 400, 500, 700, 1 k, 1.5 k, 2 k, 3 k, 4 k, 5 k, 7 k,
10 k, 15 k, 20 k, 30 k, 40 k, 50 k, 70 k, 100 k, 150 k, 200 k, 300 k, 400 k, 500 kHz
Front Panel
Description
Test Ports
Front Panel Jumpers
Probe Power
Connector
Voltage & Maximum current 1
Display
Type
Resolution
USB host port
Typical
+15 V ± 5% (400 mA)
–12.6 V ± 5% (300 mA)
General characteristic
Type –N, female, 50 Ω (nominal)
SMA, female, 50 Ω (nominal)
3 terminal connector x 2
10.4 inch TFT color LCD with touch screen
XGA (1024 x 768) 2
Universal serial bus jack, Type A configuration; female; provides con-
nection to mouse, keyboard, printer, ECal module, USB power sensor, or
USB/GPIB interface
Rear Panel
Description
Typical
General characteristic
External trigger input connector
Type
Input level
Pulse width
Polarity
BNC, female
Low threshold voltage: 0.5 V
High threshold voltage: 2.1 V
Input level range: 0 to +5 V
≥ 2 μsec
Positive or negative
External trigger output connector
Type
Maximum output current
Output level
Pulse width
Polarity
BNC, female
50 mA
Low level voltage: 0 V
High level voltage: 5 V
1 μsec
Positive or negative
External reference signal input connector
Type
Input frequency
Input level
10 MHz ± 10 ppm
–3 to +10 dBm
BNC, female
1.  Combined load for both probe connections
2.  Valid pixels are 99.99% and more. Below 0.01% (approx. 30 points) of fixed points of black, blue, green or red are not regarded as failure.