English
Language : 

E5072A Datasheet, PDF (3/23 Pages) Keysight Technologies – E5072A ENA Series Network Analyzer
3 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet
Corrected System Performance
The specifications in this section apply to measurements made with the Keysight Technologies, Inc. E5072A network analyzer
under the following conditions:
–– No averaging applied to data
–– Environmental temperature of 23 °C (± 5 °C) with less than 1 °C deviation from the
calibration temperature
–– Response and isolation calibration performed
Description
Specification
System dynamic range at test port 1, 2
(IF Bandwidth = 3 kHz)
30 to 300 kHz
65 dB (Typ.)
300 kHz to 10 MHz
82 dB
10 MHz to 6 GHz
98 dB
6 to 8.5 GHz
92 dB
(IF Bandwidth = 10 Hz)
30 to 300 kHz
90 dB (Typ.)
300 kHz to 10 MHz
107 dB
10 MHz to 6 GHz
123 dB
6 to 8.5 GHz
117 dB
Extended dynamic range at direct receiver access input 3
(IF Bandwidth = 10 Hz)
30 to 300 kHz
300 kHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
6 to 8.5 GHz
SPD
100 dB
115 dB
130 dB
128 dB
128 dB
134 dB
151 dB
147 dB
145 dB
1.  The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The effec-
tive dynamic range must take measurement uncertainty and interfering signals into account.
2.  Typical performance might not be met from 60 to 70 kHz.
3.  The direct receiver access input system dynamic range is calculated as the difference between the direct receiver access input noise floor and the
source maximum output power. The effective dynamic range must take measurement uncertainties and interfering signals into account.