English
Language : 

E4417A Datasheet, PDF (6/19 Pages) Keysight Technologies – Power Meters and E-Series E9320 Peak and Average Power Sensors
06 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors – Data Sheet
Sampling Characteristics
Sampling rate
Sampling technique
20 Msamples/second
Continuous sampling
Rear Panel Inputs/Outputs
Recorder output(s)
Remote input/output
TTL output
TTL input
Connector type
TTL output
TTL input
RS-232/422 interface
Trigger in
Trigger out
Ground
Line power
Analog 0 to 1 V, 1 kW output impedance, BNC connector.
Two outputs are available on E4417A (channels A and B).
Used to signal when measurement has exceeded a defined limit
Initiates zero and calibration cycle
RJ-45 series shielded modular jack assembly
High = 4.8 V max;
Low = 0.2 V max.
High = 3.5 V min, 5 V max;
Low = 1 V max, –0.3 V min
Serial interface for communication with an external controller.
Male plug 9-pin D-subminiature connector.
Accepts a TTL signal for initiating measurements, BNC connector
Outputs a TTL signal for synchronizing with external equipment, BNC
connector
Binding post accepts 4 mm plug or bare wire connection
Input voltage range
85 to 264 Vac, automatic selection
Input frequency range
47 to 440 Hz
Power requirement
Approximately 50 VA (14 Watts)
Remote Programming
Interface
Command language
GPIB compatibility
GPIB interface operates to IEEE 488.2 and IEC-625. RS-232 and RS-422
serial interfaces supplied as standard
SCPI standard interface commands
SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0.
Environmental Specifications
Operating environment
Temperature
Maximum humidity
Maximum altitude
0 to 55 °C
95% at 40 °C, (non-condensing)
3,000 meters (9,840 feet)
Storage conditions
Storage temperature
Non-operating maximum humidity
Non-operating maximum altitude
–20 to +70 °C
90% at 65 °C (non-condensing)
15,420 meters (50,000 feet)