04 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors – Data Sheet
Accuracy
Instrumentation
Please add the corresponding power sensor linearity percentage; see Tables 6a and 6b
for the E9320 sensors.
Average only mode
Absolute
Relative
Logarithmic
Linear
Logarithmic
Linear
± 0.02 dB
± 0.5%
± 0.04 dB
± 1.0%
Normal mode
Absolute accuracy (log)
Absolute accuracy (linear)
Relative accuracy (log)
Relative accuracy (linear)
Calibration temperature1 ± 5 °C
± 0.04 dB
± 0.8%
± 0.08 dB
± 1.6%
Temperature 0 to 55 °C
± 0.08 dB
± 1.7%
± 0.16 dB
± 3.4%
Time base accuracy
0.01%
1 mW power reference
Power output
Accuracy
Frequency
SWR
Connector type
1.00 mW (0.0 dBm). Factory set to ± 0.5% traceable to the National
Physical Laboratories (NPL), UK2
For two years
± 0.5% (23 ± 3 °C)
± 0.6% (25 ± 10 °C)
± 0.9% (0 to 55 °C)
50 MHz nominal
1.06 maximum (1.08 maximum for Option E41xA-003)
Type N (f), 50 ohms
Measurement Characteristics
Measurements
Average power
Peak power
Peak-to-average ratio
Measurements between two time offsets (time-gating)
Averaging
Averaging over 1 to 1024 readings is available for reducing noise
Measurement speed
(GPIB)
Over the GPIB, three measurement speeds are available
(normal, x 2 and fast).
The typical maximum speed is shown in the following table.
1.  Power meter is within ± 5 °C of its calibration temperature.
2.  National metrology institutes of member states of the Metre Convention, such as the National
Institute of Standards and Technology in the USA, are signatories to the ComitÈ International des