English
Language : 

E4417A Datasheet, PDF (12/19 Pages) Keysight Technologies – Power Meters and E-Series E9320 Peak and Average Power Sensors
12 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors – Data Sheet
Table 6c. Additional linearity error (normal and average only modes)
Sensor model
Temperature
( 25 ± 10 °C)
E9321A and E9325A
± 1.0%
E9322A and E9326A
± 1.0%
E9323A and E9327A
± 1.0%
Temperature
(0 to 55 °C)
± 1.0%
± 1.0%
± 1.0%
Figure 4. Typical power linearity at 25 °C for the E9323A and E9327A 5 MHz bandwidth sensors, after zero and
calibration, with associated measurement uncertainty.
Power range
Measurement
uncertainty
–30 to
–20 dBm
± 0.9%
–20 to
–10 dBm
± 0.8%
–10 to
0 dBm
± 0.65%
0 to
+10 dBm
± 0.55%
+10 to
+20 dBm
± 0.45%
+20 dBm
0 dBm
4.0 %
2%
2%
4.0 %
–60 dBm
–65 dBm
–60 dBm
–65 dBm
0 dBm
20 dBm
Figure 5. Relative mode power measurement linearity with an EPM-P series power meter, at 25 °C (typical)
Figure 5 shows the typical uncertainty in making a relative power measurement, using
the same power meter channel and the same power sensor to obtain the reference and
the measured values. It also assumes that negligible change in frequency and mismatch
error occurs when transitioning from the power level used as the reference to the power
level measured.