E4417A Datasheet, PDF (5/19 Pages) Keysight Technologies – Power Meters and E-Series E9320 Peak and Average Power Sensors
05 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors – Data Sheet
Table 2. Measurement speed for different sensor types
Sensor type
E-Series E9320 peak
and average sensors
E-Series CW and E9300
average power sensors
8480 and N8480 Series sensor
Average only mode
Normal mode3
Measurement speed (readings/second)
Normal x 2
Fast1,2
20
40
400
20
40
1000
20
40
400
20
40
N.A.
Channel functions
Storage registers
Predefined setups
A, B, A/B, B/A, A-B, B-A and Relative
10 instrument states can be saved via the Save/Recall menu.
For common wireless standards(GSM900, EDGE, NADC, iDEN, Bluetooth®,
IS-95 CDMA, W-CDMA and cdma2000®), predefined setups are provided.
Trigger
Sources
Time resolution
Delay range
Delay resolution
Hold-off
Range
Resolution
Internal, External TTL, GPIB, RS232/422
50 ns
± 1.0 s
50 ns for delays < ± 50 ms; otherwise 200 ns
1 us to 400 ms
1% of selected value (minimum of 100 ns)
Internal Trigger
Range
Level accuracy
Resolution
Latency
–20 to +20 dBm
± 0.5 dB
0.1 dB
500 ns ± 100 ns
Latency is defined as the delay between the applied RF crossing the trigger level and
the meter switching into the triggered state.
External trigger range
Trigger out
High > 2.0 V, Low < 0.8 V; BNC connector; rising or falling edge triggered;
input impedance > 1 kW
Output provides TTL compatible levels (high > 2.4 V, low < 0.4 V) and
uses a BNC connector
1.  Fast speed is not available for 8480 and N8480 series sensors.
2.  Maximum measurement speed is obtained by using binary output in free run trigger.
3.  For E9320 sensors, maximum speed is achieved using binary output in free run acquisition.