E4417A Datasheet, PDF (3/19 Pages) Keysight Technologies – Power Meters and E-Series E9320 Peak and Average Power Sensors | |||
|
◁ |
03 | Keysight | E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors â Data Sheet
Single Sensor Dynamic Range
E-series E9320 peak and average power sensors 70 dB maximum (normal mode);
85 dB maximum (average only mode)
E-series CW power sensors
90 dB
E-series E9300 average power sensors
80 dB maximum
8480 series sensors
50 dB maximum
N8480 series sensors
55 dB maximum
Display units
Absolute
Relative
Display resolution
Offset range
Video bandwidth
Watts or dBm
Percent or dB
Selectable resolution of 1.0, 0.1, 0.01, 0.001 dB
in logarithmic mode, or 1 to 4 significant digits in
linear mode.
± 100 dB in 0.001 dB increments, to compensate
for external loss or gain
5 MHz (set by meter and is sensor dependent)
Note that the video bandwidth represents the ability of the power sensor and meter to
follow the power envelope of the input signal. The power envelope of the input signal is, in
some cases, determined by the signalâs modulation bandwidth, and hence video bandwidth
is sometimes referred to as modulation bandwidth. The PCIe x16 Gen 3 link between CPU
and backplane switch provides up to 16 GB/s theoretical, 14.2 GB/s practical.
Video bandwidth/dynamic range optimization
The power measurement system, comprising the sensor and meter, has its maximum
video bandwidth defined by the E9320 sensor. To optimize the systemâs dynamic range
for peak power measurements, the video bandwidth in the meter can be set to High,
Medium and Low, as detailed in the following table. The filter video bandwidths stated in
the table are not the 3 dB bandwidths as the video bandwidths are corrected for optimal
flatness. Refer to figures 6 to 8 for information on the sensorâs peak flatness response.
A filter OFF mode is also provided.
Table 1. Video bandwidth versus peak power dynamic range
Sensor model
Video bandwidth/maximum peak power dynamic ran
OFF
High
E9321A
E9325A
300 kHz/
â40 dBm to +20 dBm
300 kHz/
â42 dBm to +20 dBm
E9322A
36 dBm to +20 dBm
1.5 MHz/
â37 dBm to +20 dBm
1.5 MHz/
â38 dBm to +20 dBm
E9323A
E9327A
5 MHz/
â32 dBm to +20 dBm
5 MHz/
â32 dBm to +20 dBm
Medium
100 kHz/
â43 dBm to +20 dBm
300 kHz/
â39 dBm to +20 dBm
1.5 MHz/
â34 dBm to +20 dBm
Low
30 kHz/
â45 dBm to +20 dBm
100 kHz/E9326A
â39 dBm to +20 dBm
300 kHz/
â36 dBm to +20 dBm
|
▷ |