English
Language : 

3000T Datasheet, PDF (34/40 Pages) Keysight Technologies – InfiniiVision 3000T X-Series Oscilloscopes
34 | Keysight | InfiniiVision 3000T X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
WaveGen – Built-in function/arbitrary waveform generator (specifications are typical) (Continued)
Frequency
Sine wave and ramp accuracy:
130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy:
[50+frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution: 0.1 Hz or 4 digits, whichever is larger
Amplitude
Range:
20 mVpp to 5 Vpp into Hi-Z 1
10 mVpp to 2.5 Vpp into 50 Ω 1
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy: 2% (frequency = 1 kHz)
DC offset
Range:
± 2.5 V into Hi-Z 1
± 1.25 V into 50 Ω 1
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy (waveform modes): ± 1.5% of offset setting ± 1% of amplitude ± 1 mV
Accuracy (DC mode): ± 1.5% of offset setting ± 3 mV
Trigger output
Trigger output available on Trig out BNC
Main output
Impedance: 50 Ω typical
Isolation: Not available, main output BNC is grounded
Protection: Overload automatically disables output
Output mode
Normal
Single-shot (arbitrary, sine, ramp, sine cardinal, exp rise/fall, cardiac, Gaussian pulse)
Digital voltmeter (Specifications are typical)
Functions
ACrms, DC, DCrms
Resolution
ACV/DCV: 3 digits
Measuring rate
100 times/second
Autoranging
Automatic adjustment of vertical amplification to maximize the dynamic range of measurements
Range meter
Graphical display of most recent measurement, plus extrema over the previous 3 seconds
Precision counter/totalizer (Specification are typical)
Counter
Source
Any analog channel or trigger qualified event
Resolution
8 digits (8 digits for trigger qualified event)
Max frequency
1 GHz
Trig qual events 1/(trigger hold off time) for trigger qualified events (max 25 MHz, minimum dead time of 40 ns)
Measurement
Frequency, period, totalize
Totalizer
Counter size
64 bit totalizing counter
Edge
Rise or fall
Gating
Positive or negative level. Select from analog channels except the source
1.  Gaussian Pulse: 4 Vpp maximum into Hi-Z; 2 Vpp maximum into 50 Ω.