English
Language : 

3000T Datasheet, PDF (28/40 Pages) Keysight Technologies – InfiniiVision 3000T X-Series Oscilloscopes
28 | Keysight | InfiniiVision 3000T X-Series Oscilloscopes - Data Sheet
Performance Characteristics
DSO and MSO 3000 X-Series oscilloscopes
3000T X-Series specification overview
3012T 3014T 3022T
3024T 3032T 3034T 3052T 3054T 3102T 3104T
Bandwidth 1 (–3 dB)
100 MHz
200 MHz
350 MHz
500 MHz
1 GHz
Calculated rise time (10 to 90%)
≤ 3.5 ns
≤ 1.75 ns
≤ 1 ns
≤ 700 ps
≤ 450 ps
Input channels
DSOX 2
MSOX 2 + 16
4
4 + 16
2
2 + 16
4
4 + 16
2
2 + 16
4
4 + 16
2
4
2
4
2 + 16 4 + 16 2 + 16 4 + 16
Maximum sample rate
5 GSa/s half channels, 2.5 GSa/s all channels
Maximum memory depth
Standard 4 Mpts, Standard segmented memory
Display size and type
8.5-inch capacitive touch gesture-enabled display
Waveform update rate
> 1,000,000 waveforms per second
Vertical system analog channels
Hardware bandwidth limits
Approximately 20 MHz (selectable)
Input coupling
AC, DC
Input impedance
Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 1.5%
Input sensitivity range
100 MHz ~ 500 MHz models: 1 mV/div to 5 V/div2 (1 MΩ and 50 Ω)
1 GHz models: 1 mV/div to 5 V/div2 (1 MΩ), 1 mV/div to 1 V/div (50 Ω)
Vertical resolution
8 bits (measurement resolution is 12 bits with averaging)
Maximum input voltage
300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
With N2843A 10:1 probe: 300 Vrms
Frequency de-rating (assumes sine wave input): 400 Vpk until 40 kHz. Then de-rates at 20 db/dec until 6 Vpk
DC vertical accuracy
± [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] 2
DC vertical gain accuracy 1
± 2.0% full scale 2
DC vertical offset accuracy
± 0.1 div ± 2 mV ± 1% of offset setting
Channel-to-channel isolation
> 100:1 from DC to maximum specified bandwidth of each model (measured with same V/div and coupling on
channels)
Offset range
± 2 V (1 mV/div to 200 mV/div)
± 50 V (> 200 mV/div to 5 V/div)
Vertical system digital channels
Digital input channels
16 digital (D0 to D15. pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Thresholds
Threshold per pod
Threshold selections
TTL (+1.4 V), 5 V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
User-defined threshold range
± 8.0 V in 10 mV steps
Maximum input voltage
± 40 V peak CAT I; transient overvoltage 800 Vpk
Threshold accuracy1
± (100 mV + 3% of threshold setting)
Maximum input dynamic range
± 10 V about threshold
Minimum voltage swing
500 mVpp
Input impedance
100 kΩ ± 2% at probe tip
Input capacitance
~8 pF
Vertical resolution
1 bit
1.  Denotes warranted specifications, all others are typical.
2.  Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration temperature. 1 mV/div and 2 mV/div are a magnification of
4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV div and 2 mV/div sensitivity setting.