3000T Datasheet, PDF (30/40 Pages) Keysight Technologies – InfiniiVision 3000T X-Series Oscilloscopes
30 | Keysight | InfiniiVision 3000T X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
Trigger system
Trigger sources
Trigger modes
Trigger coupling
Trigger holdoff range
Trigger sensitivity
Internal 1
External 1
Trigger level range
Any channel
External
Trigger type selections
Zone (HW zone qualifier)
Edge
Edge then edge (B trigger)
Pulse width
Runt
Setup and hold
Rise/fall time
Analog channel (1 ~ 4), digital channel (D0 ~ D15), line, external, WaveGen (1 or mod) (FM/FSK)
Normal (triggered): Requires trigger event for scope to trigger
Auto: Triggers automatically in absence of trigger event
Single: Triggers only once on a trigger event, press [Single] again for scope to find another trigger event, or
press [Run] to trigger continuously in either Auto or Normal mode
Force: front panel button that forces a trigger
DC: DC coupled trigger
AC: AC coupled trigger, cutoff frequency: < 10 Hz (internal); <50 Hz (external)
HF reject: High frequency reject, cutoff frequency ~ 50 kHz
LF reject: Low frequency reject, cutoff frequency ~ 50 kHz
Noise reject: Selectable OFF or ON, decreases sensitivity 2x
40 ns to 10.00 s
< 10 mV/div: Greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
200 mVpp from DC to 100 MHz
350 mVpp 100 MHz to 200 MHz
± 6 div from center screen
±8V
Trigger on user-defined zones drawn on the display. Applies to one analog channel at a time. Specify zones as
either “must intersect” or “must not intersect.” Up to two zones. > 200,000 scans/sec update rate
Supported modes: normal, peak detect, high resolution
Also works simultaneously with the serial trigger and mask/limit test
Trigger on a rising, falling, alternating or either edge of any source
Arm on a selected edge, wait a specified time, then trigger on a specified count of another selected edge
Trigger on a pulse on a selected channel, whose time duration is less than a value, greater than a value, or inside
a time range
Minimum duration setting: 2 ns (500 MHz, 1 GHz), 4 ns (350 MHz), 6 ns (200 MHz), 10 ns (100 MHz)
Maximum duration setting: 10 s
Range minimum: 10 ns
Trigger on a position runt pulse that fails to exceed a high level threshold. Trigger on a negative runt pulse that
fails to exceed a low level threshold. Trigger on either polarity runt pulse based on two threshold settings. Runt
triggering can also be time-qualified (< or >) with a minimum time setting of 2 ~ 10 ns and maximum timesetting
of 10 s
Minimum time setting: 2 ns (500 MHz, 1 GHz), 4 ns (350 MHz), 6 ns (200 MHz)
10 ns (100 MHz)
Trigger and clock/data setup and/or hold time violation. Setup time can be set from –7 to 10 s. Hold time can be
set from 0 s to 10 ns
Trigger on rise-time or fall-time edge speed violations (< or >) based on user-selectable threshold
Select from (< or >) and time settings range between
Minimum: 1 ns (500 MHz, 1 GHz), 2 ns (350 MHz), 3 ns (200 MHz), 5 ns (100 MHz)
Maximum: 10 s
1.  Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration