English
Language : 

3000T Datasheet, PDF (32/40 Pages) Keysight Technologies – InfiniiVision 3000T X-Series Oscilloscopes
32 | Keysight | InfiniiVision 3000T X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
Waveform measurements
Cursors 2
Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
Dual cursor accuracy: ± [DC vertical gain accuracy + 0.5% full scale] 1
Units: Seconds(s), Hz (1/s), phase (degrees), ratio (%)
Automatic measurements Measurements continuously updated with statistics. Cursors track last selected measurement. Select up to eight
measurements from the list below:
Snapshot All: Measure all single waveform measurements (31)
Voltage: Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot, average- N cycles, average- full
screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles, AC RMS- full screen (std deviation), ratio- N cycle,
ratio- full screen
Time: Period, frequency, counter, + width, - width, burst width, +duty cycle, -duty cycle, bit rate, rise time, fall time, delay,
phase, X at min Y, X at max Y
Count: Positive pulse count, negative pulse count, rising edge count, falling edge count
Mixed: Area- N cycles, area- full screen
Automatic measurement Available via BenchVue
logging
Counter (see pg 32 for Built-in frequency counter
8-digit precision counter) Source: On any analog or digital channel
Resolution: 5 digits
Maximum frequency: Bandwidth of scope
Waveform math
Number of math functions
Two, displays FFT and one math simultaneously. Can be cascaded
Arithmetic
Add, subtract, multiply, divide, differentiate, integrate, FFT, Ax + B, squared, square root, absolute value, common
logarithm, natural logarithm, exponential, base 10 exponential, low pass filter, high pass filter, averaged value,
smoothing, envelope, magnify, max hold, min hold, measurement trend, chart logic bus (Timing or State)
Enhanced FFT
Record size Up to 64 kpts resolution
Window types Hanning, Flat Top, Rectangular, Blackman-Harris
Time gated FFT Gate the time range of data for FFT analysis in the zoom view. For time and frequency domain correlated analysis.
Waveforms FFT, max hold, min hold, average
Peak search Max 11 peaks, threshold and excursion control
Search, navigate, and lister
Type
Edge, pulse width, rise/fall, runt, frequency peak, serial bus 1, serial bus 2
Copy
Copy to trigger, copy from trigger
Frequency peak Source
Math functions
Max # of peaks 11
Control
Results order in frequency or amplitude
Result display
Event lister or navigation. Manual or auto scroll via navigation or touch event lister entry to jump to a specific event
Display characteristics
Display
8.5-inch capacitive touch/gesture enabled TFT LCD
Resolution
800 (H) x 480 (V) pixel format (screen area)
Graticules
8 vertical divisions by 10 horizontal divisions with intensity controls
Format
YT, XY, and Roll
Maximum waveform update rate
> 1,000,000 wfms/s
Persistence
Off, infinite, variable persistence (100 ms to 60 s)
Intensity gradation
64 intensity levels
1.  Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration temperature.
2.  1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity
setting.