English
Language : 

4000 Datasheet, PDF (30/37 Pages) C&D Technologies – Toroidal Surface Mount Inductors
30 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Dual-channel WaveGen built-in function/arbitary waveform generator (specifications are typical)
Sine
Frequency range
0.1 Hz to 20 MHz
Amplitude flatness
± 0.5 dB (relative to 1 kHz)
Harmonic distortion
–40 dBc
Spurious (non harmonics)
–40 dBc
Total harmonic distortion
1%
SNR (50-Ω load, 500-MHz BW)
40 dB (Vpp ≥ 0.1 V); 30 dB (Vpp < 0.1 V)
Square wave/pulse
Frequency range
0.1 Hz to 20 MHz
Duty cycle
20 to 80%
Duty cycle resolution
Larger of 1% or 10 ns
Rise/fall time
19 ns (10 to 90%)
Overshoot
< 2%
Asymmetry (at 50% DC)
± 1% ± 5ns
Jitter (TIE RMS)
500 ps
Ramp/triangle wave
Frequency range
0.1 Hz to 200 kHz
Linearity
1%
Variable symmetry
0 to 100%
Symmetry resolution
1%
Pulse
Frequency range
0.1 Hz to 10 MHz
Pulse width
20 ns minimum
Pulse with resolution
10 ns
Edge time
Fixed at 19 ns (not variable)
Overshoot
< 2%
Noise
Bandwidth
20 MHz typical
Sine cardinal (Sinc)
Frequency range
0.1 Hz to 1.0 MHz
Exponential rise/fall
Frequency range
0.1 Hz to 5.0 MHz
Cardiac
Frequency range
0.1 Hz to 200.0 kHz
Gaussian pulse
Frequency range
0.1 Hz to 5.0 MHz
Arbitrary
Waveform length
1 to 8,192 points
Amplitude resolution
10 bits (including sign bit) 2
Repetition rate
0.1 Hz to 12 MHz
Sample rate
100 MSa/s
Filter bandwidth
20 MHz
Frequency
Sine wave and ramp accuracy
130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy
[50 + frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution
0.1 Hz or 4 digits, whichever is larger
Amplitude
Range: Minimum
20 mVpp if |offset| ≤ 0.5 Vpp into Hi-Z 1
10 mVpp if |offset| ≤ 0.5 Vpp into 50 Ω 1
Range: Maximum
10 Vpp except, 9 Vpp if Sinc or Cardiac, 7.5 Vpp if Gaussian
pulse into Hi-Z; 5 Vpp/4.5 Vpp into 50 Ω
Resolution
100 μV or 3 digits, whichever is higher
Accuracy
1.5% (frequency = 1 kHz)
1.  Sinc, cardiac and Gaussian pulse: ± 1.25 V into Hi-Z; ± 625 mV into 50 Ω.
2.  Full resolution is not available at output due to internal attenuator stepping.