English
Language : 

5989-8853EN Datasheet, PDF (25/37 Pages) Keysight Technologies – A guide to power product solutions to match your test and measurement needs
25 | Keysight | Power Products – October 2015 - Selection Guide
USB Modular Source Measure Unit
Source and measure DC voltage/current reliably
The Keysight USB modular source mea-
sure unit (SMU) allows you to perform
sweeps and make measurements using a
single device. The SMU offers voltage and
current programming/readback with high
accuracy measurement capabilities. You
can conigure each of the three channels
separately or in a matrix—in series or par-
allel—for increased power. It comes bun-
dled with Keysight Measurement Manager
(AMM) software that includes a command
logger function to help you convert SCPI
commands into snippets of VEE, V, C+ and
C# code.
– Three-channel, four-quadrant opera-
tion (± 20 V, ± 120 mA)
– High measurement sensitivity of
100 pA with 16-bit resolution
– 0.1% basic accuracy
– Low current measurement
capability down to nA levels
– Embedded test script able to
support three channels with
coherent source and
measurement capabilities (for
U2723A)
– IV Curve application support in the
Keysight Measurement
Manager Software (for U2723A)
– Faster rise/fall time (for U2723A)
– Hi-Speed USB 2.0 (480 Mbps)
U2722A
Model
U2722A/23A
Number of outputs
3
Output ratings (at 0 °C to 50 °C)
Voltage
–20 V to 20 V per channel
Current
–120 mA to 120 mA per channel
Model
Voltage
programming/
readback
Current
programming/
readback
Range
±2V
± 20 V
± 1 µA
± 10 µA
± 100 µA
± 1 mA
± 10 mA
± 120 mA
U2722A/23A
Accuracy 1
0.075% + 1.5 mV
Resolution
0.1 mV
0.05% + 10 mV
1 mV
0.085% + 0.85 nA
0.085% + 8.5 nA
0.075% + 75 nA
0.075% + 750 nA
100 pA
1 nA
10 nA
100 nA
0.075% + 7.5 µA
0.1% + 100 µA
1 µA
20 µA
Model
U2722A
U2723A
Rise/fall time (ms) 1
For resistive
measurement 2
Range
± 1 µA
± 10 µA
± 100 µA
± 1 mA
± 10 mA
± 120 mA
Accuracy 1
170.0
18.0
6.0
1.0
1.0
1.0
Accuracy 1
15.0
5.0
1.0
1.0
1.0
1.0
1 Drive 50% of 1 V or 10 V output with a resistive load. Rise time is from 10% to
90% of program voltage change at maximum current. Fall time is from 90% to
10% of program voltage change at maximum current.
2 Measurements obtained are per default bandwidth setting.