08 | Keysight | E4982A LCR Meter - Data Sheet
Definition of each parameter (continued)
Zs =
Mode 2
Oscillator level= 1 dBm, Average factor ≥ 8 ± (13 + 0.5 × F) [mΩ]
Oscillator level= 1 dBm, Average factor < 8 ± (15 + 0.5 × F) [mΩ]
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (16 + 0.5 × F) [mΩ]
factor ≥ 8
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (24 + 0.5 × F) [mΩ]
factor < 8
–33 dBm ≤ Oscillator level< -20 dBm, Aver- ±(24+0.5×F) [mΩ]
age factor ≥ 8
–33 dBm ≤ Oscillator level < –20 dBm, Aver- ± (64 + 0.5 × F) [mΩ]
age factor < 8
Oscillator level < –33 dBm
± (133 + 0.5 × F) [mΩ]
Mode 3
Oscillator level = 1 dBm, Average factor ≥ 8 ± (12 + 0.5 × F) [mΩ]
Oscillator level = 1 dBm, Average factor < 8 ± (14 + 0.5 × F) [mΩ]
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (15 + 0.5 × F) [mΩ]
factor ≥ 8
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (20 + 0.5 × F) [mΩ]
factor < 8
–33 dBm ≤ Oscillator level < –20 dBm, Aver- ± (20 + 0.5 × F) [mΩ]
age factor ≥ 8
–33 dBm ≤ Oscillator level < –20 dBm, Aver- ± (50 + 0.5 × F) [mΩ]
age factor < 8
Oscillator level < –33 dBm
± (100 + 0.5 × F) [mΩ]
Yo =
Within 23 ± 5 °C from the calibration temperature. Measurement accuracy applies when the calibration is performed at
23 ± 5 °C. When the calibration is performed beyond 23 ± 5 °C, the measurement accuracy decreases to half that de-
scribed. (F: Frequency [MHz])
Measurement Time:
Mode 1
Oscillator level = 1 dBm, Average factor ≥ 8 ± (22 + 0.15 × F) [µS]
Oscillator level = 1 dBm, Average factor < 8 ± (28 + 0.15 × F) [µS]
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (30 + 0.15 × F) [µS]
factor ≥ 8
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (53 + 0.15 × F) [µS]
factor < 8
–33 dBm ≤ Oscillator level < –20 dBm, Aver- ± (52 + 0.15 × F) [µS]
age factor ≥ 8
–33 dBm ≤ Oscillator level < –20 dBm, Aver- ± (110 + 0.15 × F) [µS]
age factor < 8
Oscillator level < –33 dBm
± (247 + 0.15 × F) [µS]
Mode 2
Oscillator level = 1 dBm, Average factor ≥ 8 ± (20 + 0.15 × F) [µS]
Oscillator level = 1 dBm, Average factor < 8 ± (23 + 0.15 × F) [µS]
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (24 + 0.15 × F) [µS]
factor ≥ 8
–20 dBm ≤ Oscillator level < 1 dBm, Average ± (35 + 0.15 × F) [µS]
factor < 8
–33 dBm ≤ Oscillator level< –20 dBm, Aver- ± (35 + 0.15 × F) [µS]
age factor ≥ 8
–33 dBm ≤ Oscillator level< –20 dBm, Aver- ± (63 + 0.15 × F) [µS]
age factor < 8
Oscillator level < –33 dBm
± (133 + 0.15 × F) [µS]