English
Language : 

E4982A Datasheet, PDF (19/22 Pages) Agilent(Hewlett-Packard) – New Standard for High-Speed Component Tests
19 | Keysight | E4982A LCR Meter - Data Sheet
Analyzer Environmental Specifications and Dimensions
Operating environment
Temperature
Error-corrected temperature range
Humidity
Altitude
Vibration
+5 °C to +40 °C
23 °C (± 5 °C) with < 5 °C deviation from calibration temperature
20% to 80% at wet bulb temperature < +29 °C (non-condensation)
0 to 2,000 m (0 to 6,561 feet)
0.21 G maximum, 5 Hz to 500 Hz
Non-operating environment
Temperature
–10 °C to +60 °C
Humidity
20% to 90% at wet bulb temperature < 40 °C (non-condensation)
Altitude
0 to 4,572 m (0 to 15,000 feet)
Vibration
2.1 G maximum, 5 Hz to 500 Hz
Dimensions, weight
Weight
Main unit: 13 kg, test head: 250 g with plate
Figure 12. Front view