English
Language : 

E4982A Datasheet, PDF (17/22 Pages) Agilent(Hewlett-Packard) – New Standard for High-Speed Component Tests
17 | Keysight | E4982A LCR Meter - Data Sheet
Line power
Frequency
Voltage
VA max
47 to 63 Hz
90-264 VAC (Vpeak > 120 V)
300 VA max.
EMC, safety, environment and compliance
EMC
European Council Directive 2004/108/EC
IEC 61326-1:2012
EN 61326-1:2013
CISPR 11:2009 +A1:2010
EN 55011: 2009 +A1:2010
Group 1, Class A
IEC 61000-4-2:2008
EN 61000-4-2:2009
4 kV CD / 8 kV AD
IEC 61000-4-3:2006 +A1:2007 +A2:2010
EN 61000-4-3:2006 +A1:2008 +A2:2010
3 V/m, 80-1000 MHz, 1.4 - 2.0 GHz / 1V/m, 2.0 - 2.7 GHz, 80% AM
IEC 61000-4-4:2004 +A1:2010
EN 61000-4-4:2004 +A1:2010
1 kV power lines / 0.5 kV signal lines
IEC 61000-4-5:2005
EN 61000-4-5:2006
0.5 kV line-line / 1 kV line-ground
IEC 61000-4-6:2008
EN 61000-4-6:2009
3 V, 0.15-80 MHz, 80% AM
IEC 61000-4-8:2009
EN 61000-4-8:2010
30A/m, 50/60Hz
IEC 61000-4-11:2004
EN 61000-4-11:2004
0.5-300 cycle, 0% / 70%
NOTE-1:
When tested at 3 V/m according to EN61000-4-3, the measurement accuracy will be within
specifications over the full immunity test frequency range except when the analyzer frequency is
identical to the transmitted interference signal test frequency.
NOTE-2:
When tested at 3 V according to EN61000-4-6, the measurement accuracy will be within
specifications over the full immunity test frequency range except when the analyzer frequency is
identical to the transmitted interference signal test frequency.
ICES-001:2006 Group 1, Class A
AS/NZS CISPR11:2004
Group 1, Class A
KN11, KN61000-6-1 and KN61000-6-2
Group 1, Class A