03 | Keysight | E4982A LCR Meter - Data Sheet
Basic Measurement Characteristic
Measurement parameters
Impedance parameters
|Z|, |Y|, Ls, Lp, Cs, Cp, Rs, Rp, X, G, B, D, Q, Ãz [°], Ãz [rad], Ãy [°], Ãy [rad],
User defined parameter (A maximum of four parameters can be displayed at one time.)
Measurement range
Impedance parameters
140 m⦠to 4.8 kâ¦
(Frequency = 1 MHz, Averaging factor = 8, Measurement time mode = 3,
Oscillator level = 1 dBm, Measurement uncertainty ⤠± 10%, Calibration is performed within
23 °C ± 5 °C, Measurement is performed within ± 5 °C from the calibration temperature)
Source Characteristics
Frequency
Range
Resolution
Uncertainty
1 MHz to 300 MHz (Option 030)
1 MHz to 500 MHz (Option 050)
1 MHz to 1 GHz (Option 100)
1 MHz to 3 GHz (Option 300)
100 kHz
± 10 ppm (23 °C ± 5 °C)
± 20 ppm (5 °C to 40 °C)
Oscillator level
Cable Length = 1m:
Power range (When 50 ⦠LOAD is
connected to test port)
Current range (When SHORT is
connected to test port)
Voltage range (When OPEN is
connected to test port)
Uncertainty (When 50 ⦠LOAD is
connected to test port)
Resolution
â40 dBm to 1dBm
0.0894 mArms to 10 mArms
4.47 mVrms to 502 mVrms
(23 °C ± 5 °C)
± 2 dB (frequency ⤠1 GHz)
± 3 dB (frequency > 1 GHz)
(5 °C to 40 °C)
± 4 dB (frequency ⤠1 GHz)
± 5 dB (frequency > 1 GHz)
0.1 dB (When the unit is set at mV or mA, the entered value is rounded to 0.1 dB resolution.)
Cable Length = 2m (When option 002 is used):
Power range
Subtract the following attenuation from the power (setting value) at 1 m cable length:
Attenuation [dB] = 0.42 âf (f: Frequency [GHz])
Uncertainty (When 50 ⦠LOAD is
connected to test port)
(23 °C ± 5 °C)
± 3 dB (frequency ⤠1 GHz)
± 4 dB (frequency > 1 GHz)
(5 °C to 40 °C)
± 5 dB (frequency ⤠1 GHz)
± 6 dB (frequency > 1 GHz)
Resolution
0.1 dB (When the unit is set at mV or mA, the entered value is rounded to 0.1 dB resolution.)
Output impedance
Output impedance
50 ⦠(nominal)
|
▷ |