English
Language : 

U3606B Datasheet, PDF (7/20 Pages) Keysight Technologies – Multimeter I DC Power Supply
07 | Keysight | U3606B Multimeter I DC Power Supply - Data Sheet
Digital multimeter specifications (continued)
AC specifications
Table 2. AC accuracy specifications ± (% of reading + % of range)
Function
Range1
Frequency range
1 year
23 ºC ± 5 ºC
Temperature coefficient 0 to
18 ºC
28 to 55 ºC
True rms AC voltage2
100.000 mV3
1.00000 V
to 750.00 V5, 6
20 to 45 Hz
45 Hz to 10 kHz
10 to 30 kHz
30 to 100 kHz4
20 to 45 Hz7
45 Hz to 10 kHz
10 to 30 kHz
30 to 100 kHz4, 8
1 + 0.1
0.2 + 0.1
1.5 + 0.3
5 + 0.3
1 + 0.1
0.2 + 0.1
1 + 0.1
3 + 0.2
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
True rms AC current
10.0000 mA
to 3.0000 A9
20 to 45 Hz
45 Hz to 1 kHz
1 to 10 kHz
1.5 + 0.110
0.5 + 0.1
2 + 0.211
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
1. 20% over range on all ranges, except for 750 Vac range.
2. Specifications stated are for input signals greater than 5% of range except for the 100 mV range. No square-wave output are to be
used as the signal output.
3. 100 mV range: specifications stated are for input signals greater than 10% of range.
4. Additional error 0.003% of full scale per kHz to be added when signal input changes less than 10% of range.
5. Available ranges: 1.00000 V, 10.0000 V, 100.000 V, 750.00 V.
6. For 750 V range: 847 V is readable.
7. For 750 V range: the accuracy is specified for input less than 200 Vrms.
8. For 750 V range: the accuracy is specified for input less than 300 Vrms.
9. Available ranges: 10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A.
10. For 3 A range: the accuracy is specified for input less than 3 A.
11. For 1 A and 3 A ranges: the accuracy is specified for frequencies less than 5 kHz.
The specification of the AC+DC measurement will be the sum of the AC and DC accuracy. The frequency range will be
from 50 Hz for 5½ digit resolution and 225 Hz for 4½ digit resolution.