English
Language : 

U3606B Datasheet, PDF (11/20 Pages) Keysight Technologies – Multimeter I DC Power Supply
11 | Keysight | U3606B Multimeter I DC Power Supply - Data Sheet
Digital multimeter specifications (continued)
Supplemental characteristics (continued)
AC current
Measurement method
AC coupled true rms
Maximum input current
10 mA to 3.0 A DC or AC rms1
Burden voltage and shunt resistance
– < 0.2 V, 10 Ω for 10 mA range
– < 0.2 V, 1 Ω for 100 mA range
– < 0.3 V, 0.05 Ω for 1 A range
– < 0.7 V, 0.05 Ω for 3 A range
Input protection
Protected with 3.15 A/500 V, FF fuse
Crest factor
For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at
full scale.
Peak input
300% of range. Limited by maximum input.
Response time
Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested
input signal at the same range.
1. Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a smaller current or
offset current measurement immediately after a high current measurement, ensure that the U3606B is cooled down.
Resistance
Measurement method
Open circuit voltage
Input protection
Response time
Two-wire, open-circuit voltage limited to < 5 V
< +5.0 Vdc
1000 Vrms on all ranges, < 0.3 A short circuit
Approximately 0.15 seconds for 1 MΩ and ranges below 1 MΩ
Low-resistance
Measurement method
Input protection
Open circuit voltage
Four-wire, the test current is sent from the FORCE terminals and resistance measured at the
SENSE terminals.
– FORCE terminals: Protected with a 3.15 A/250 V FF fuse
– SENSE terminals: 1000 Vrms on all ranges, < 0.3 A short circuit
< +8.6 VDC
Continuity
Measurement method
Open circuit voltage
Audible tone
Input protection
0.83 mA ± 0.2% constant current source
< +5.0 Vdc
Continuous beeping when reading is less than the threshold resistance of 10 Ω at 1.0 kΩ
range