13 | Keysight | U3606B Multimeter I DC Power Supply - Data Sheet
DC power supply specifications
Safety considerations
The U3606B is a safety class I instrument, which means it has a protective earth terminal. The terminal must be connected
to an earth ground through a power source with a 3-wire ground receptacle.
The DC power supply performance specifications are listed in the following pages. Specifications are warranted in the
temperature range of 0 to 55 °C with a fix resistive load. Supplemental characteristics — which are not warranted, but are
descriptions of performance — are determined either by design or testing.
Specification assumptions:
– Specifications stated are after 60-minutes of warm-up and with no load
– Operating temperature at 18 to 28 °C (64.4 to 82.4 °F)
– Accuracy is expressed as ± (% of output + offset) at 23 °C ± 5 °C
– Temperature coefficient: Add [0.1 × (the specified accuracy) / °C] for 0 to 18 °C and 28 to 55 °C
– Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for 30 to 55 °C
Performance specifications
Table 8. DC power supply performance specifications
Parameter
Output ratings
Programming accuracy
1 year (@ 23 ºC ± 5 ºC),
±(% of output + offset)
Readback accuracy
1 year over GPIB and USB or
front panel with respect to
actual output (@ 23 ºC ± 5 ºC),
±(% of output + offset)
Ripple and noise
With outputs ungrounded, or
with either output terminal
grounded, 20 Hz to 1 MHz
Voltage
Current
Voltage
Current
Specifications
S1S2
S1
AUTO
30 V/1 A
0.05% +
5 mV
0.05% +
5 mV
0.15% +
3 mA
0.15% +
3 mA
0.05% +
5 mV
0.05% +
5 mV
0.15% +
3 mA
0.15% +
3 mA
Normal mode
voltage
Normal mode
current
S1m
100 mA/30 V
0.05% +
5 mV
0.05% +
0.15 mA
0.05% +
5 mV
0.15% +
0.15 mA
S2
8 V/3 A
0.05% +
5 mV
0.15% +
3 mA
0.05% +
5 mV
0.15% +
3 mA
< 2 mVrms; < 30 mVpp
< 1 mArms
S2m
1000 mV/3 A
0.05% +
0.5 mV
0.15% +
3 mA
0.05% +
0.5 mV
0.15% +
3 mA
(continued on next page)