DC power supply specifications (continued)
Performance specifications (continued)
Table 8. DC power supply performance specifications
Specifications
Parameter
S1S2
S1
S1m
S2
S2m
Output ratings
AUTO
30 V/1 A
100 mA/30 V 8 V/3 A
1000 mV/3 A
Front terminal load regulation1
±(% of output + offset)
Voltage
Current
< 3 mV +
(6 mV/A)
< 0.03% +
0.3 mA
< 3 mV +
(6 mV/A)
< 0.03% +
0.3 mA
< 3 mV +
(6 mV/A)
< 0.03% +
0.03 mA
< 3 mV +
(6 mV/A)
< 0.03% +
0.3 mA
< 0.3 mV +
(6 mV/A)
< 0.03% +
0.3 mA
Rear terminal load regulation
±(% of output + offset)
Voltage
Current
< 0.01% +
3 mV
< 0.03% +
0.3 mA
< 0.01% +
3 mV
< 0.03% +
0.3 mA
< 0.01% +
3 mV
< 0.03% +
0.03 mA
< 0.01% +
3 mV
< 0.03% +
0.3 mA
< 0.01% +
0.3 mV
< 0.03% +
0.3 mA
Line regulation
Voltage
Current
3 mV typical
1.5 mA
typical
3 mV typical
1.5 mA
typical
3 mV typical
0.15 mA
typical
3 mV typical
1.5 mA
typical
0.3 mV typical
1.5 mA
typical
Programming resolution
Voltage
Current
1 mV
0.1 mA
1 mV
0.1 mA
1 mV
0.01 mA
1 mV
0.1 mA
0.1 mV
0.1 mA
Readback resolution
Voltage
Current
1 mV
0.1 mA
1 mV
0.1 mA
1 mV
0.01 mA
1 mV
0.1 mA
0.1 mV
0.1 mA
Front panel resolution
Voltage
Current
1 mV
0.1 mA
1 mV
0.1 mA
1 mV
0.01 mA
1 mV
0.1 mA
0.1 mV
0.1 mA
Transient response time
Less than 100 µs for output to recover to within 15 mV following a change in
output current from full load to half load or vice versa
Command processing time
Average time for output voltage to begin to change after receipt of digital data
when instrument is connected directly to the USB or GPIB is less than 100 ms
Over-voltage protection
(for CC mode)
Accuracy: 0.5% + 0.5 V
Activation time2: < 2 ms
Over-current protection
(for CV mode)
Accuracy: 0.5% + 0.05 A
Activation time2: < 2 ms
1. The terminal sense is related to the resistance of the contacts or leads, and proportional to the load condition.
2. Average time for the detection of OVP or OCP condition. The output will be dropped down and set to standby within 20 ms.