U3606B Datasheet, PDF (5/20 Pages) Keysight Technologies – Multimeter I DC Power Supply
05 | Keysight | U3606B Multimeter I DC Power Supply - Data Sheet
Digital multimeter specifications
Specification assumptions:
– Specifications stated are after 60 minutes of warm-up and for 5½-digit resolution
– One-year calibration cycle, with calibration temperature of 23 °C ± 2 °C
– Operating temperature: 18 to 28 °C (64.4 to 82.4 °F)
– Accuracy is expressed as ± (% of reading + % of range)
– Temperature coefficient: Add [0.1 x (the applicable accuracy) / °C] for 0 to 18 °C and 28 to 55 °C
– Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for 30 to 55 °C
DC specifications
Table 1. DC accuracy specifications ± (% of reading + % of range)
Function
Range1
Resolution
Test current
or burden
voltage
24 hours2
23 ºC ± 1 ºC
90 days
23 ºC ± 5 ºC
1 year
23 ºC ± 5 ºC
Temperature
coefficient
0 to 18 ºC
28 to 55 ºC
19.9999 mV 0.1 μV
—
100.000 mV 1 μV
—
DC voltage
1.00000 V
10.0000 V
10 μV
100 μV
—
—
100.000 V 1 mV
—
1000.00 V 10 mV
—
0.012 + 0.04
0.012 + 0.008
0.012 + 0.005
0.012 + 0.005
0.012 + 0.005
0.012 + 0.005
0.015 + 0.04
0.015 + 0.008
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.015 + 0.005
0.025 + 0.04
0.025 + 0.008
0.025 + 0.005
0.025 + 0.005
0.025 + 0.005
0.025 + 0.005
0.0015 + 0.0040
0.0015 + 0.0008
0.0010 + 0.0005
0.0020 + 0.0005
0.0015 + 0.0005
0.0015 + 0.0005
DC current3
10.0000 mA
100.000 mA
1.00000 A
3.0000 A
0.1 μA
1 μA
10 μA
100 μA
< 0.2 V
< 0.2 V
< 0.3 V
< 0.7 V
0.05 + 0.015
0.05 + 0.005
0.05 + 0.007
0.05 + 0.007
0.05 + 0.015
0.05 + 0.005
0.05 + 0.007
0.05 + 0.007
0.05 + 0.015
0.05 + 0.005
0.15 + 0.007
0.15 + 0.007
0.0060 + 0.0005
0.0060 + 0.0005
0.0100 + 0.0005
0.0150 + 0.0010
Resistance4
100.000 Ω
1000.00 Ω
10.0000 kΩ
100.000 kΩ
1.00000 MΩ
10.0000 MΩ
100.000 MΩ
1 mΩ
10 mΩ
100 mΩ
1Ω
10 Ω
100 Ω
1 kΩ
0.83 mA
0.83 mA
100 µA
10 µA
900 nA
205 nA
205 nA ||
10 MΩ
0.04 + 0.008
0.04 + 0.005
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.20 + 0.005
1.60 + 0.005
0.04 + 0.008
0.04 + 0.005
0.04 + 0.005
0.04 + 0.005
0.05 + 0.005
0.20 + 0.005
1.60 + 0.005
0.05 + 0.008
0.05 + 0.005
0.05 + 0.005
0.05 + 0.005
0.06 + 0.005
0.25 + 0.005
2.00 + 0.005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0050 + 0.0005
0.0150 + 0.0005
0.1500 + 0.0005
1. 20% over range on all ranges, except for 20 mVdc, 1000 Vdc, and 3 Adc range.
2. Relative to calibration standards.
3. Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a lower current or offset
current immediately after a high current measurement, ensure that the U3606B has cooled down.
4. Specifications stated are for 2-wire resistance measurements using Null math operation. Without Null, add a 0.2 Ω error. To
eliminate the noise interference, which might be induced due to the test leads, a shielded test cable is recommended for measuring
resistance above 100 kΩ.
(continued on next page)