English
Language : 

N1045A Datasheet, PDF (23/37 Pages) Keysight Technologies – Wide-Bandwidth Oscilloscope Mainframe and Modules
23 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Clock recovery
The DCA platform offers clock recovery capability in the form of DCA plug-in modules such as the 83496B or standalone clock
recovery solutions such as the N1076A/N1077A. Specifications for 83496B modules are summarized below. For more information on
standalone clock recovery solutions, please download the N1076A/N1077A data sheet from www.keysight.com/find/N1077A.
Clock recovery modules
Channel type
Data rates (divide by 2 for clock signals)
Minimum input level to acquire lock
(voltage or OMA 1)
Output random jitter (RMS) 2
Clock recovery adjustable loop
bandwidth range (user selectable)
Loop bandwidth accuracy
Tracking range
Aquisition range
Internal splitter ratio
Input return loss
Input insertion loss
See footnotes on next page.
83496B-100
83496B-101
Differential or single-ended electrical Single-mode or multimode optical, differential or single-ended
electrical (no internal electrical splitters)
Standard: 50 Mb/s to 7.1 Gb/s continuous tuning
Option 200: 50 Mb/s to 14.2 Gb/s continuous tuning
Option 201: 7.1 to 14.2 Gb/s continuous tuning
For rates above 14 Gb/s, the N1076A and N1077A operate from 50 Mb/s to 32 Gb/s
150 m Vpp
Single-mode (OMA 1):
–11 dBm to 50 Mb/s to 11.4 Gb/s
–8 dBm to > 11.4 Gb/s
–12 dBm to 7.1 to 14.2 Gb/s (with Option 200)
–14 dBm to 1 to 7.1 Gb/s
–15 dBm to 50 Mb/s to 1 Gb/s
Multimode 1310 nm (OMA 1):
–10 dBm to 50 Mb/s to 11.4 Gb/s
–7 dBm to > 11.4 Gb/s
–11 dBm to 7.1 to 14.2 Gb/s (with Option 200)
–13 dBm to 1 to 7.1 Gb/s
–14 dBm to 50 Mb/s to 1 Gb/s
Multimode 850 nm (OMA 1):
–8 dBm to 50 Mb/s to 11.4 Gb/s
–7 dBm to > 11.4 Gb/s
–9 dBm to 7.1 to 14.2 Gb/s (with Option 200)
–11 dBm to 1 to 7.1 Gb/s
–12 dBm to 50 Mb/s to 1 Gb/s
Electrical: 150 m Vpp
Internal recovered clock trigger:
< 500 fs 7.2 to 11.4 Gb/s (300 fs to 10 Gb/s)
< 700 fs 4.2 to 7.2 Gb/s, 11.4 Gb/s to 14.2 Gb/s (400 fs to 4.25 Gb/s, 500 fs to 2.5 Gb/s)
< 3 mUI 50 Mb/s to 4.2 Gb/s (700 fs to 1.25 Gb/s)
Front panel recovered clock:
< 700 fs 7.2 to 11.4 Gb/s (300 fs to 10 Gb/s)
< 900 fs 4.2 to 7.2 Gb/s, 11.4 to 14.2 Gb/s (400 fs to 4.25 Gb/s, 500 fs to 2.5 Gb/s)
< 4 mUI 50 Mb/s to 4.2 Gb/s (700 fs to 1.25 Gb/s)
Standard: 270 kHz or 1.5 MHz 3
Option 300: 15 kHz to 10 MHz 4 continuous tuning (fixed value or a constant rate/N ratio)
Standard: ± 30%
Option 300: ± 25% for transition density = 0.5 and data rate 155 Mb/s to 11.4 Gb/s
(± 30% for 0.25 ≤ transition density ≤ 1.0 and all data rates)
± 2500 ppm, 83496B
± 1000 ppm, 83496A
± 5000 ppm
50/50
50/50 single-mode
30/70 multimode
Electrical signals have input only (no internal power dividers)
22 dB (DC to 12 GHz) electrical
20 dB single-mode, 16 dB multimode
16 dB (12 to 20 GHz) electrical
22 dB min (DC to 12 GHz) electrical
16 dB min (12 to 20 GHz) electrical
7.2 dB max (DC to 12 GHz) electrical
4 dB max single-mode optical
7.8 dB max (12 to 20 GHz) electrical
4 dB max multimode optical (no electrical data output signal path)