N1045A Datasheet, PDF (15/37 Pages) Keysight Technologies – Wide-Bandwidth Oscilloscope Mainframe and Modules
15 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Single-mode and multimode optical/electrical (Continued)
Multiple and single-mode
86105C
86105D
86105D
optical/electrical modules
Option 281
Optical channel specifications (Continued)
User calibrated accuracy
Single-mode
± 3% ± 200 nW ± 2% ± 100 nW ± 3% ± 200 nW
± Power meter
± Power meter
± Power meter
Uncertainty
Uncertainty
Uncertainty
< 5 °C change
Multimode (characteristic)
± 10% ± 200 nW ± 10% ± 200 nW ± 10% ± 200 nW
± Power meter
± Power meter
± Power meter
Uncertainty
Uncertainty
Uncertainty
< 5 °C change
Maximum input power
Maximum non-destruct average
0.5 mW (–3 dBm) 5 mW (7 dBm)
5 mW (+7 dBm)
1310/1550 nm
Single mode:
6 mW (+8 dBm)
850 nm
Multimode:
3 mW (+5 dBm)
Maximum non-destruct peak
5 mW (+7 dBm) 10 mW (10 dBm) 10 mW (+10 dBm)
Fiber input
62.5/125 μm
62.5/125 μm
62.5/125 µm
User-selectable
Connector
Polarization dependent loss at
0.2 dB
1550 nm
(characteristic)
Input return loss (HMS-10 connector 850 nm > 13 dB 27 dB single-mode > 24 dB (1550 nm
fully filled fiber)
single mode)
1310 nm/1550 nm 14 dB multimode > 24 dB (1310 nm
single mode)
> 24 dB
> 13 dB (850 nm
multimode)
86115D
86115D
Option 002, 102, 142 1 Option 004, 104, 144 1
± 2% ± 100 nW
± Power meter
Uncertainty
< 5 °C change
± 10% ± 200 nW
± Power meter
Uncertainty
± 2% ± 100 nW
± Power meter
Uncertainty
± 10% ± 200 nW
± Power meter
Uncertainty
5 mW (7 dBm)
5 mW (7 dBm)
10 mW (10 dBm)
62.5/125 μm
User-selectable
Connector
10 mW (10 dBm)
62.5/125 μm
User-selectable
Connector
27 dB single-mode
14 dB multimode
27 dB single-mode
14 dB multimode
1.  Options 102 and 104 do not include filters for 14.025 Gb/s. Options 142 and 144 include filters for 14.025 Gb/s and no other rates.