English
Language : 

2000 Datasheet, PDF (20/24 Pages) Make-Ps – Single, Dual, Triple Output DC/DC Converter
20 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet
InfiniiVision X-Series Physical Characteristics
Connectivity
Standard ports
One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol
Two USB 2.0 hi-speed host ports, front and rear panel
Supports memory devices, printers and keyboards
Optional ports
GPIB, LAN, WVGA video out
General and environmental characteristics
Power line consumption
100 W
Power voltage range
100 to 120 V, 50/60/400 Hz; 100 to 240 V, 50/60 Hz ± 10% auto ranging
Temperature
Operating: 0 to +55 °C
Non-operating: –30 to +71 °C
Humidity
Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C
Altitude
Operating: Up to 4,000 m, Non-operating 15,300 m
Electromagnetic compatibility Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
Safety
UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Dimensions (W x H x D)
381 mm (15 in) x 204 mm (8 in) x 142 mm (5.6 in)
Weight
Net: 3.9 kg (8.5 lbs), shipping: 4.1 kg (9.0 lbs)
Nonvolatile storage
Reference waveform display
2 internal waveforms or USB thumb drive
Waveform storage
Set up, .bmp, .png, .csv, ASCII, XY, reference waveforms, .alb, .bin, lister, mask , HDFS
Max USB flash drive size
Supports industry standard flash drives
Set ups without USB flash drive 10 internal setups
Set ups with USB flash drive
Limited by size of USB drive
Included standard with oscilloscope
Standard 5-year warranty 1 (90 days for unserialized accessories such as passive probes)
Standard secure erase
Standard probe
N2862B 150 MHz 10:1 passive probe
Standard one per channel for 70 and 100 MHz models
N2863B 300 MHz, 10:1 passive probe
Standard one per channel for 200 MHz models
N6459-60001 8-channel logic probe and accessory kit
Standard on MSO models or with DSOX2MSO upgrade
Built-in help language support for English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian, Portuguese and
Italian, Certificate of Calibration, Documentation CD
Interface language support GUI menus: English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian,
Portuguese, Thai, Polish and Italian
Localized power cord
1.  Applies to all orders on or after 1/1/2013.
For MET/CAL procedures, click on the Cal Labs solutions link below Cal Labs Solutions http://www.callabsolutions.com/products/Keysight/
These procedures are FREE to customers.