English
Language : 

2000 Datasheet, PDF (16/24 Pages) Make-Ps – Single, Dual, Triple Output DC/DC Converter
16 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
Acquisition modes
Normal
Peak detect
Averaging
High resolution mode
Segmented
Trigger system
Trigger modes
Trigger coupling
Trigger source
Trigger sensitivity (internal) 1
Trigger sensitivity (external) 1
External trigger input
Trigger type selections
Edge
Pulse width
Pattern
Video
I²C (optional)
SPI (optional)
CAN (optional)
LIN (optional)
RS232/422/485/UART
(optional)
Capture glitch as narrow as 500 ps at all timebase settings
Select from 2, 4, 8, 16, 64... to 65,536
12 bits of resolution when ≥ 20 µs/div
Re-arm time= 19 µs (minimum time between trigger events)
–– Normal (triggered): requires trigger event for scope to trigger
–– Auto: triggers automatically in absence of trigger event
–– Single: triggers only once on a trigger event, press [Single] again for scope to find another trigger event, or press
[Run] to trigger continuously in either Auto or Normal mode
–– Force: front panel button that forces a trigger
Coupling selections: AC, DC, noise reject, LF reject and HF reject
Each analog channel, each digital channel (MSO models or DSOX2MSO upgrade, Ext, WaveGen, line)
< 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
200 mV (DC to 100 MHz); 350 mV (100 MHz - 200 MHz)
Included on all models
All 2000 X-Series models
Trigger on a rising, falling, alternating or either edge of any source
Trigger on a pulse on a selected channel, whose time duration is less than a value, greater than a value, or inside a time
range
–– Minimum duration setting: 2 to 10 ns (depends on bandwidth)
–– Maximum duration setting: 10 s
Trigger when a specified pattern of high, low, and don’t care levels on any combination of analog, digital, or trigger
channels is [entered | exited]. Pattern must have stabilized for a minimum of 2 ns to qualify as a valid trigger condition.
Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL,
SECAM, PAM-M)
Trigger on I²C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame with address and/or data
values. Also trigger on missing acknowledge, address with no accq, restart, EEPROM read, and 10-bit write.
Trigger on SPI (Serial Peripherial Interface) data pattern during a specific framing period. Supports positive and negative
Chip Select framing as well as clock Idle framing and userspecified number of bits per frame.
Trigger on CAN (controller area network) version 2.0A and 2.0B signals. Trigger on the start of frame (SOF) bit (standard).
Remote frame ID (RTR), data frame ID (~RTR), remote or data frame ID, data frame ID and data, error frame, all errors,
acknowledge error and overload frame.
Trigger on LIN (Local Interconnect Network) sync break, sync frame ID, or frame ID and data
Trigger on Rx or Tx start bit, stop bit or data content
1.  Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and from ± 10 °C firmware calibration
temperature.