Vertical gain
The âVertical Gainâ calibration process allows writing
4 gain calibration variables to the device ï¬ash mem-
ory. The Vpp measurement utility is used to acquire
the correct gain setting.
Rev B1, Page 12/22
4. Press the âCal. GL1â button
5. Press âWrite Cal. to iC227 ï¬ash memoryâ button
Figure 29: Write calibration data
6. Press the scale @50 mV/div âSet for Calâ button
and follow the instructions.
Figure 27: Vertical gain
Vertical gain calibration steps
1. Make sure that scope was powered on for mini-
mum 10 minutes
2. Open Calibration - Edit Calibration Data
3. Press the scale @200 mV/div âSet for Calâ button
and follow the instructions.
Figure 30: Amplitude calibration
7. Press the âCal. GH 1â button
8. Press âWrite Cal. To iC227 ï¬ash memoryâ button
Figure 28: Amplitude calibration
Horizontal Calibration
The âHorizontalâ calibration process allows writing 2
calibration variables to the device ï¬ash memory.
Figure 31: Horizontal calibration
|
▷ |