Menu - File:
Load Setup
Save Setup to a simple text ï¬le e.g. conï¬g.txt
Exit Program
Figure 23: File menu
CALIBRATION
The scope iC227 does not have a certiï¬cate of cali-
bration, but this device has software and hardware de-
signed for auto time base calibration and calibration of
vertical gains and offsets on both channels. There is
also a software feature which allows manual calibra-
tion of the time base in case that the accuracy of the
internal crystal and auto calibration are not sufï¬cient.
If the device is to be used in production where certiï¬ed
calibration is required, then there must be done peri-
odically calibration against a known calibrated source.
Vertical offset calibration steps
1. Make sure that scope was powered on for mini-
mum 10 minutes
2. Open Calibration - Edit Calibration Data
3. Press the âSet for Cal Lâ button and follow the in-
structions
Vertical offset
The âVertical Offsetâ calibration process allows writing
4 offset calibration variables to the device ï¬ash mem-
ory. This will adjust the channel trace to be at the level
of the channel pointer âCH1>â and âCH2>â seen on
the left side of the screen when the input voltage is
0.00 V. There are 2 calibration registers for each chan-
nel: 500 mV/div and 50 mV/div. The ï¬rst one is respon-
sible for the vertical resolutions 1 V-0.5 V-0.2 V, while
the second represents 100 mV-50 mV-20 mV-10 mV.
Figure 25: Vertical calibration
4. Press the âCal. Offset Lâ button
5. Press âWrite Cal. to iC227 ï¬ash memoryâ button
Figure 24: Vertical offset
Figure 26: Write calibration data
6. Press the âSet for Cal Hâ button and follow the
instructions
7. Press the âCal. Offset Hâ button
8. Press âWrite Cal. to iC227 ï¬ash memoryâ button
|
▷ |