UNIT 3
CERTIFICATION METER
CERTIFICATION METER - Creating User-Definable Cabling Standards
Some companies have their own set of optical fiber loss parameters – such as fiber loss, connector loss and
splice loss – that they need their network to adhere to. Thus, the FiberMeter allows the user to configure two
user-definable cabling standards for the purpose of fiber certification.
These standards are defined when selecting the fiber standard during the Link Wizard process.
NOTE: one of the key parameters for defining these custom cabling standards is fiber loss (dB per kilometer).
Each custom standard supports a different range of optical losses (for up to 2 wavelengths), thus it is important
to know what the fiber loss is because this determines which user-definable standard to use.
USER DEFINED #1 supports fiber losses for up to two wavelengths from 0.01 dB to 2.55 dB per kilometer
USER DEFINED #2 supports fiber losses for up to two wavelengths from 0.1 dB to 25.5 dB per kilometer
1)
Press F1 to scroll through the list of fiber standards.
FIBER STANDARDS
2)
When the correct custom standard is highlighted, press F3 to EDIT.
USER DEFINED #1
USER DEFINED #2
TIA-568B 3/CAN-T529
ISO/IEC 11801
_____________________
NEXT SELECT EDIT
3)
From the SELECT WAVELENGTHS screen, press F1 to scroll to the SELECT WAVELENGTHS(2)
first wavelength to configure.
*1 850nm
*2 980nm
4)
Press F2 to select this wavelength as the first wavelength.
1300nm
1310nm
_____________________
NEXT SEL1 SEL2
5)
Enter the loss per kilometer in dB for 62.5/125 MM fiber in the entry field. 850nm06205um0MM
Decimals can be entered by holding the 0 key to scroll through the loss/km(0001-2055dB):
.-/
special characters. If it is not necessary to enter a value, leave the entry [010000]
field blank.
_____________________
6)
Press DONE to continue. Repeat Steps 5 & 6 for the remaining fiber types. <--- SHIFT --->
7)
Enter the loss per connection in dB, then press DONE to continue.
Enter Loss/Connector
(dB):
[010000]
8)
Enter the loss per splice in dB, then press DONE to continue.
_____________________
<--- SHIFT --->
Enter Loss/Splice
(dB):
[010000]
9)
Highlight the other wavelength to use for this standard and press F3 to
select.
_____________________
<--- SHIFT --->
10) Repeat steps 5 through 8 for the second wavelength. Press DONE to return
to the fiber standard selection screen, then press F2 to select the custom standard.
3-15