English
Language : 

SBAU202 Datasheet, PDF (27/47 Pages) Texas Instruments – Performance Demonstration Kit for the ADS130E08
www.ti.com
ADS130E08 ANALYSIS TOOLS
5.3 FFT Tool
The FFT subtab displays the data in the frequency domain by performing an FFT on the eight channels.
Details of the FFT, including SNR, THD, and more, can be viewed using the FFT Analysis button located
at the bottom of the display. Figure 27 shows the FFT display.
Figure 27. Analysis→FFT Graph of Normal Electrode Configuration
5.3.1 Coherent Frequency Calculator
The red box labeled 1 in Figure 27 shows the Coherent Frequency Calculator. Coherent sampling in an
FFT is defined in Equation 3:
fAIN
fSAMPLE
=
NWINDOW
NTOTAL
where:
• fAIN = input frequency
• fSAMPLE = ADS130E08 sampling frequency
• NWINDOW = number of odd integer cycles during a given sampling period
• NTOTAL = number of data points (in power of 2) that is used to create the FFT.
(3)
If the conditions for coherent sampling can be met, the FFT results for a periodic signal are optimized. The
ideal AIN frequency is a value that is calculated based on the sampling rate so that the coherent sampling
criteria can be met.
SBAU202 – June 2012
Submit Documentation Feedback
Performance Demonstration Kit for the ADS130E08
27
Copyright © 2012, Texas Instruments Incorporated