U1453A Datasheet, PDF (9/20 Pages) Keysight Technologies – Insulation Resistance Testers
09 | Keysight | U1450A/U1460A Insulation Resistance Testers - Data Sheet
U1451A/U1452A/U1452AT earth bond resistance speciications
Earth-bond resistance specifications with accuracy of ± (% of reading + number of least significant digit)1, 2
Range
60 Ω
600 Ω
6 kΩ
60 kΩ
Resolution
0.01 Ω
0.1 Ω
0.001 kΩ
0.01 kΩ
Accuracy
U1451A
1.5% + 3
1.5% + 3
1.5% + 3
1.5% + 3
U1452A/U1452AT
1.0% +3
1.0% +3
1.0% +3
1.0% +3
Open circuit voltage
> 4 V and < 7 V
1. The following statements are true for earth-bond resistance tests:
– Overload protection: < 2 V and 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse
– Short circuit: > 200.0 mA as resistance ≤ 2 Ω
2. The accuracy is speciied after the Null function is used to subtract the test lead resistance and thermal effect (by shorting the test leads)
U1453A/U1461A earth bond resistance speciications
Earth-bond resistance specifications with accuracy of ± (% of reading + number of least significant digit)1
Range
6 Ω2
60 Ω2
600 Ω2
6 kΩ
60 kΩ
Resolution
0.001 Ω
0.01 Ω
0.1 Ω
0.001 kΩ
0.01 kΩ
Accuracy
0.5% + 20
0.5% + 2
0.5% + 2
0.5% + 2
0.5% + 2
Open circuit voltage
> 4 V and < 7 V
1. The following statements are true for earth-bond resistance tests:
– Overload protection: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse
– Short circuit: > 200.0 mA as resistance ≤ 2 Ω
2. The accuracy of the 6 to 600 Ω range is speciied after the Null function is used to subtract the test lead resistance and thermal effect (by shorting
the test leads)