08 | Keysight | U1450A/U1460A Insulation Resistance Testers - Data Sheet
U1453A/U1461A insulation resistance speciications
Insulation resistance specifications with accuracy of ± (% of reading + number of least significant digit)
Test voltage
50 V
100 V
250 V
500 V
1000 V
Range
6 MΩ
< 50 MΩ
~60 GΩ
6 MΩ
60 MΩ
< 100 MΩ
~60 GΩ
6 MΩ
60 MΩ
< 250 MΩ
~200 GΩ
6 MΩ
60 MΩ
< 500 MΩ
~200 GΩ
6 MΩ
60 MΩ
600 MΩ
< 1 GΩ
~200 GΩ
Resolution
0.001 MΩ
0.01 MΩ
~0.01 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
~0.01 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
~0.1 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
~0.1 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
0.001 GΩ
~0.1 GΩ
Accuracy
1.5% + 5
1.5% + 5
1.5% + 52
1.5% + 5
1.5% + 5
1.5% + 5
1.5% + 52
1.5% + 5
1.5% + 5
1.5% + 5
1.5% + 52
1.2% + 5
1.2% + 5
1.2% + 5
1.2% + 52
1.2% + 5
1.2% + 5
1.2% + 5
1.2% + 5
1.2% + 52
Test current
1 mA @ 50 kΩ
1 mA @ 100 kΩ
1 mA @ 250 kΩ
1 mA @ 500 kΩ
1 mA @ 1 MΩ
1. The voltage indication on the display refers to the voltage at the DUT (device under test), and the accuracy is according to the DC voltage
measurement. The marked test voltage and the actual test voltage may be different if the test voltage is adjusted in the Setup. Refer to the table
below for more details.
Ω
Mega
position
(Mark)
Default test voltage
(factory)
Deviation
Adjustable test voltage
(user)
Incremental
1000 V
500 V
250 V
100 V
50 V
1000 V
500 V
250 V
100 V
50 V
2.0 V
1.5 V
0.2%
0.3%
10 to 1100 V 10 to 600 V
1.5 V
0.6%
10 to 300 V
1.5 V
1.5%
10 to 120 V
1.0 V
2.0%
10 to 60 V
1V
1V
1V
1V
1V
2. Additional accuracy is to be added to the basic accuracy as shown in the table below.
Voltage
Above
Basic accuracy
Additional accuracy
1000 V
1 GΩ
1.2% + 5
0.05%/GΩ
500 V
500 MΩ
1.2% + 5
0.1%/GΩ
250 V
250 MΩ
1.5% + 5
0.2%/GΩ
100 V
100 MΩ
1.5% + 5
0.5%/GΩ
50 V
50 MΩ
1.5% + 5
1.0%/GΩ