Speciications
U1451A/U1452A/U1452AT insulation resistance speciications
Insulation resistance specifications with accuracy of ± (% of reading + number of least significant digit)
Test voltage
50 V
100 V
250 V
500 V
1000 V
Range
6 MΩ
< 50 MΩ
~60 GΩ
6 MΩ
60 MΩ
< 100 MΩ
~60 GΩ
6 MΩ
60 MΩ
< 250 MΩ
~200 GΩ
6 MΩ
60 MΩ
< 500 MΩ
~200 GΩ
6 MΩ
60 MΩ
600 MΩ
< 1 GΩ
~200 GΩ
Resolution
0.001 MΩ
0.01 MΩ
~0.01 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
~0.01 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
~0.1 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
~0.1 GΩ
0.001 MΩ
0.01 MΩ
0.1 MΩ
0.001 GΩ
~0.1 GΩ
Accuracy
2% + 5
2% + 5
2% + 52
2% + 5
2% + 5
2% + 5
2% + 52
1.5% + 5
1.5% + 5
1.5% + 5
1.5% + 52
1.5% + 5
1.5% + 5
1.5% + 5
1.5% + 52
1.5% + 5
1.5% + 5
1.5% + 5
1.5% + 5
1.5% + 52
Test current
1 mA @ 50 kΩ
1 mA @ 100 kΩ
1 mA @ 250 kΩ
1 mA @ 500 kΩ
1 mA @ 1 MΩ
1. The voltage indication on the display refers to the voltage at the DUT (device under test), and the accuracy is according to the DC voltage mea-
surement. The default test voltage is shown in the table below.
Ω
Mega
position
(Mark)
Test voltage
Test accuracy
1000 V
500 V
250 V
100 V
50 V
1000 V
500 V
250 V
100 V
50 V
0% to +20 % 0% to +20 % 0% to +20 % 0% to +20 % 0% to +20 %
The test voltage across a resistor of a value of UN × (1000 Ω/V) shall not differ by more than 10% relative to noload (open) voltage, as a result of
possibly present AC voltage components in the output voltage, when a capacitor of 1 µF is connected in parallel with the insulation resistance to
be measured. UN represents the nominal output test voltage.
2. Additional accuracy is to be added to the basic accuracy as shown in the table below.
Voltage
Above
Basic accuracy
Additional accuracy
1000 V
1 GΩ
1.5% + 5
0.1%/GΩ
500 V
500 MΩ
1.5% + 5
0.2%/GΩ
250 V
250 MΩ
1.5% + 5
0.4%/GΩ
100 V
100 MΩ
2.0% + 5
1.0%/GΩ
50 V
50 MΩ
2.0% + 5
2.0%/GΩ