English
Language : 

N1923A Datasheet, PDF (9/19 Pages) Keysight Technologies – Wideband Power Sensors
09 | Keysight | 8990B Peak Power Analyzer and N1923A/N1924A Wideband Power Sensors - Data Sheet
Product Characteristics
The following specifications are applicable only when the N1923A/N1924A wideband power sensors are used with the 8990B peak
power analyzer. Using the 8990B with other supported sensors might yield different results.
Power requirements
Operating
Non-operating conditions
Dimensions (W x D x H)
Weight
Sound pressure level
Electromagnetic
compatibility
Safety
100 to 120 V (at 50/60/400 Hz)
100 to 240 V (at 50 to 60 Hz)
Maximum power dissipated at 375 W
Operating temperature from 5 to 40 °C
Relative humidity up to 95% at 40 °C (non-condensing)
Operating altitude up to 4600m (15000 ft.)
Operating random vibration at 5 to 500 Hz, 10 minutes/axis, 0.21 g (rms)
Non-operating temperature from –40 to +70 °C
Relative humidity up to 90% at 65 °C
Non-operating altitude up to 4600 m (15000 ft.)
Non-operating random vibration at 5 to 500 Hz, 10 minutes/axis, 2.09 g (rms); resonant search at 5 to 500 Hz, swept
sine, 1 octave/minute, sweep rate at 0.5 g (0 peak), 5 minutes resonant, dwell at 4 resonance/axis
430 mm (16.9 in) x 347 mm (13.7 in) x 330 mm (13.0 in)
< 16 kg (net)
< 23.5 kg (shipping)
45 dB
Complies with the essential requirements of the European (EC) Directives as follows:
–– IEC 61326-2-1:2005/EN 61326-2-1:2006
–– CISPR 11:2003/EN 55011:2007 (Group 1, Class A)
The product also meets the following EMC standards:
–– Canada: ICES-1:2004
–– Australia/New Zealand: AS/NZS CISPR 11:2004
Conforms to the following product specifications:
–– EN61010-1: 2001/IEC 61010-1:2001
–– CAN/CSA C22.2 No. 61010-1-04
–– ANSI/UL std No. 61010-1-2004