English
Language : 

N1923A Datasheet, PDF (13/19 Pages) Keysight Technologies – Wideband Power Sensors
13 | Keysight | 8990B Peak Power Analyzer and N1923A/N1924A Wideband Power Sensors - Data Sheet
N1923A/N1924A Wideband Power Sensor Specifications
Sensor model Frequency range Dynamic range
N1923A
50 MHz to 18 GHz –35 to +20 dBm
N1924A
50 MHz to 40 GHz –35 to +20 dBm
Rise/fall time
≤ 3 ns (applicable for
frequencies of ≥ 500 MHz)
≤ 3 ns (applicable for
frequencies of ≥ 500 MHz)
Damage level
+23 dBm (average power);
+30 dBm (< 1 μs duration, peak power)
+23 dBm (average power);
+30 dBm (< 1 μs duration, peak power)
Connector type
Type N (m)
2.4 mm (m)
The N1921A/N1922A P-series wideband power sensors are compatible for use with the 8990B peak power analyzer.
Maximum SWR
Frequency band
50 MHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 26.5 GHz
26.5 GHz to 40 GHz
N1923A
1.2
1.26
N1924A
1.2
1.26
1.3
1.5
Sensor calibration uncertainty 1
Frequency band
50 MHz to 500 MHz
500 MHz to 1 GHz
1 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 26.5 GHz
26.5 GHz to 40 GHz
Physical characteristics
Dimensions
Weights with cable
Fixed sensor cable lengths
N1923A
4.5%
4.0%
4.0%
5.0%
N1923A
N1924A
Option 105
Option 106
Option 105
Option 106
N1924A
4.3%
4.2%
4.4%
4.7%
5.9%
6.0%
135 mm x 40 mm x 27 mm (5.3 in x 1.6 in x 1.1 in)
127 mm x 40 mm x 27 mm (5.0 in x 1.6 in x 1.1 in)
0.4 kg (0.88 Ib)
0.6 kg (1.32 Ib)
1.5 m (5-feet)
3.0 m (10-feet)
Environmental conditions
General
Operating
Temperature
Maximum humidity
Maximum altitude
Storage
Non-operating storage temperature
Non-operating maximum humidity
Non-operating maximum altitude
Complies with the requirements of the EMC Directive 89/336/EEC
0 to 55 °C
95% at 40 °C (non-condensing)
3,000 meters (9,840 feet)
–30 to +70 °C
90% at 65 °C (non-condescending)
15,420 meters (50,000 feet)
1.  Beyond 70% humidity, an additional 0.6% should be added to these values.